Medical proton dosimetry using radioluminescence from aluminium oxide crystals attached to optical-fiber cables
-
Andersen, C. E. (författare)
-
- Medin, Joakim (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
Medin, J. (författare)
-
visa fler...
-
- Grusell, Erik (författare)
- Uppsala universitet,Avdelningen för sjukhusfysik
-
Jain, M. (författare)
-
- Mattsson, Sören (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2007
- 2007
- Engelska.
-
Ingår i: Nuclear Instruments and Methods in Physics Research Section A. - : Elsevier BV. - 0168-9002 .- 1872-9576 .- 0167-5087. ; 580:1, s. 466-468
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
https://lup.lub.lu.s...
-
visa färre...
Abstract
Ämnesord
Stäng
- The prime objective of this study is to investigate if radioluminescence (RL) from carbon-doped aluminum oxide (Al2O3:C) crystals potentially can be used for absorbed dose-rate measurements during proton radiotherapy. The RL from two separate (2 mg) Al2O3:C crystals attached to optical-fiber cables were recorded during irradiations in water in a 175 MeV clinical proton beam. The RL response for low-LET protons in the plateau region of the Bragg curve was found to closely resemble that observed for a clinical 6 MV X-ray beam. An identical response was found in the Bragg peak (where the dose-averaged LET is about 4 keV/mu m) for absorbed doses less than 0.3 Gy. For doses in the range of 0.3-3Gy, we observed a significant decrease in luminescence efficiency with LET. At 3 Gy, the luminescence efficiency was about 60% in the Bragg-peak region. The study implies that the RL-signal from Al2O3:C could potentially be suitable for medical proton dosimetry in the 0-0.3 Gy range even without any LET-dependent correction factors.
Ämnesord
- NATURVETENSKAP -- Fysik -- Acceleratorfysik och instrumentering (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Accelerator Physics and Instrumentation (hsv//eng)
Nyckelord
- radiation-induced luminescence
- dosimetry
- neutron and proton
- MEDICINE
- MEDICIN
- radiation-induced luminescence
- dosimetry
- neutron and proton
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas