Novel sensitive cardiac troponin I immunoassay free from troponin I-specific autoantibody interference
-
Savukoski, Tanja (författare)
-
Jacobino, Jenna (författare)
-
Laitinen, Paivi (författare)
-
visa fler...
-
- Lindahl, Bertil (författare)
- Uppsala universitet,Kardiologi,Uppsala kliniska forskningscentrum (UCR)
-
- Venge, Per (författare)
- Uppsala universitet,Biokemisk struktur och funktion
-
Ristiniemi, Noora (författare)
-
Wittfooth, Saara (författare)
-
Pettersson, Kim (författare)
-
visa färre...
-
(creator_code:org_t)
- Walter de Gruyter GmbH, 2014
- 2014
- Engelska.
-
Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 52:7, s. 1041-1048
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Background: Cardiac troponins (cTnI and cTnT) are the recommended biomarkers of myocardial infarction. As cTn-specific autoantibodies (cTnAAb) can interfere with the cTn detection by state-of-the-art cTnI assays, our objective was to develop a sensitive cTnI immunoassay free from this analytical interference. Methods: The assay used antibody-coated spots containing three capture Mabs/Fabs directed against the N-terminus, midfragment and C-terminus of cTnI and a europium chelate-labeled tracer Mab against the C-terminus. Following a 3-h sample incubation and washing, cTnI was quantified by time-resolved fluorometry. Results: The limit of detection (LoD) was 2.9 ng/L and the assay was linear up to 50,000 ng/L. The total precision of 10% CV was not reached, but 20% CV was reached at 10 ng/L. Mean cTnI (10-50,000 ng/L) recoveries were 100% and 119% in three cTnAAb-positive and two cTnAAb-negative individuals, respectively, verifying the interference resistance of the antibody design used. On average, Architect hs-cTnI assay gave seven-fold higher cTnI concentrations than the new assay but the correlation between the assays was good (r=0.958). Of apparently healthy individuals (n=159), 18% had measurable cTnI values (>LoD) and 10% were cTnAAb-positive. The proportion of measurable cTnI values, however, was significantly higher in cTnAAb-positive individuals (13/16, median cTnI 8.5 ng/L) than in cTnAAb-negative individuals (15/143, median cTnI
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)
Nyckelord
- antibody-coated spot
- autoantibody
- cardiac troponin I
- immunoassay
- interference
- myocardial infarction