Automatic extraction of forward stroke volume using dynamic PET/CT : a dual-tracer and dual-scanner validation in patients with heart valve disease.
-
- Harms, Hendrik Johannes (författare)
- Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
-
- Tolbod, Lars Poulsen (författare)
- Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
-
- Hansson, Nils Henrik Stubkjær (författare)
- Aarhus Univ Hosp, Dept Cardiol, DK-8200 Aarhus N, Denmark
-
visa fler...
-
- Kero, Tanja (författare)
- Uppsala universitet,Radiologi
-
- Örndahl, Lovisa Holm (författare)
- Uppsala universitet,Kardiologi
-
- Kim, Won Yong (författare)
- Aarhus Univ Hosp, Dept Cardiol, DK-8200 Aarhus N, Denmark
-
- Bjerner, Tomas (författare)
- Uppsala universitet,Radiologi
-
- Bouchelouche, Kirsten (författare)
- Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
-
- Wiggers, Henrik (författare)
- Aarhus Univ Hosp, Dept Cardiol, DK-8200 Aarhus N, Denmark
-
- Frøkiær, Jørgen (författare)
- Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
-
- Sörensen, Jens (författare)
- Uppsala universitet,Radiologi,Department of Nuclear Medicine & PET Centre, Aarhus University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2015-10-26
- 2015
- Engelska.
-
Ingår i: EJNMMI Physics. - : Springer Science and Business Media LLC. - 2197-7364. ; 2:1
- Relaterad länk:
-
https://doi.org/10.1...
-
visa fler...
-
https://uu.diva-port... (primary) (Raw object)
-
https://ejnmmiphys.s...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- BACKGROUND: The aim of this study was to develop and validate an automated method for extracting forward stroke volume (FSV) using indicator dilution theory directly from dynamic positron emission tomography (PET) studies for two different tracers and scanners.METHODS: 35 subjects underwent a dynamic (11)C-acetate PET scan on a Siemens Biograph TruePoint-64 PET/CT (scanner I). In addition, 10 subjects underwent both dynamic (15)O-water PET and (11)C-acetate PET scans on a GE Discovery-ST PET/CT (scanner II). The left ventricular (LV)-aortic time-activity curve (TAC) was extracted automatically from PET data using cluster analysis. The first-pass peak was isolated by automatic extrapolation of the downslope of the TAC. FSV was calculated as the injected dose divided by the product of heart rate and the area under the curve of the first-pass peak. Gold standard FSV was measured using phase-contrast cardiovascular magnetic resonance (CMR).RESULTS: FSVPET correlated highly with FSVCMR (r = 0.87, slope = 0.90 for scanner I, r = 0.87, slope = 1.65, and r = 0.85, slope = 1.69 for scanner II for (15)O-water and (11)C-acetate, respectively) although a systematic bias was observed for both scanners (p < 0.001 for all). FSV based on (11)C-acetate and (15)O-water correlated highly (r = 0.99, slope = 1.03) with no significant difference between FSV estimates (p = 0.14).CONCLUSIONS: FSV can be obtained automatically using dynamic PET/CT and cluster analysis. Results are almost identical for (11)C-acetate and (15)O-water. A scanner-dependent bias was observed, and a scanner calibration factor is required for multi-scanner studies. Generalization of the method to other tracers and scanners requires further validation.
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas
- Av författaren/redakt...
-
Harms, Hendrik J ...
-
Tolbod, Lars Pou ...
-
Hansson, Nils He ...
-
Kero, Tanja
-
Örndahl, Lovisa ...
-
Kim, Won Yong
-
visa fler...
-
Bjerner, Tomas
-
Bouchelouche, Ki ...
-
Wiggers, Henrik
-
Frøkiær, Jørgen
-
Sörensen, Jens
-
visa färre...
- Artiklar i publikationen
-
EJNMMI Physics
- Av lärosätet
-
Uppsala universitet