The HemoScreen, a novel haematology analyser for the point of care
-
- Ben-Yosef, Yaara (författare)
- PixCell Med Technol Ltd, Yokneam Ilit, Israel
-
- Marom, Barak (författare)
- PixCell Med Technol Ltd, Yokneam Ilit, Israel
-
- Hirshberg, Galit (författare)
- PixCell Med Technol Ltd, Yokneam Ilit, Israel
-
visa fler...
-
- D'Souza, Carol (författare)
- Univ Westminster, Fac Sci & Technol, Biomed Sci, London, England
-
- Larsson, Anders (författare)
- Uppsala universitet,Biokemisk struktur och funktion
-
- Bransky, Avishay (författare)
- PixCell Med Technol Ltd, Yokneam Ilit, Israel
-
visa färre...
-
(creator_code:org_t)
- 2016-01-19
- 2016
- Engelska.
-
Ingår i: Journal of Clinical Pathology. - : BMJ. - 0021-9746 .- 1472-4146. ; 69:8, s. 720-725
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- BACKGROUND AND AIMS: A haematology analyser, based on a new technology, is presented herein. The analyser that provides a complete blood count (CBC) and five-part differential accepts disposable cartridges containing all required reagents, making it maintenance-free and ideal for point-of-care (POC) settings. The test reproducibility and imperviousness to analytical errors are attributed to the imaging-based analysis employed. Imaging enables cell-morphology-based differentiation, which is analogous to the gold standard microscopic analysis. This article presents the HemoScreen new technology and evaluates its performance through a small-scale study conducted in its designated clinical settings.METHODS: Thirty anticoagulated whole blood samples were analysed on the HemoScreen and Sysmex XE-2100. Linear regression was performed for the methods comparison. Two samples with 15 replicates were processed for imprecision. Ease of use of the device was also considered.RESULTS: The HemoScreen demonstrated acceptable imprecision and good agreement with the Sysmex XE-2100. The white blood cells (WBCs), red blood cells (RBCs), haemoglobin (HGB), haematocrit (HCT), platelets (PLT), neutrophils, lymphocytes and eosinophils have coefficients of correlation (r) >0.97. For mean cell volume (MCV), mean cell HGB (MCH) and RBC distribution width (RDW), r values ranged from 0.92 to 0.96. For mean cell HGB concentration (MCHC) and monocytes r=0.82 was demonstrated. User-friendliness and suitability of the device for operation in the designated POC settings was also confirmed.CONCLUSIONS: The HemoScreen employs innovative technologies of viscoelastic focusing and microfluidics within a disposable cartridge for an image-based blood cell analysis. By providing accurate and repeatable CBC and five-part differential results within minutes and maintaining the simplicity of operation, the HemoScreen could have far-reaching implications for use at POC. Further extended evaluation is in progress.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Hematologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Hematology (hsv//eng)