Sökning: onr:"swepub:oai:DiVA.org:uu-320922" >
A setup for element...
A setup for element specific magnetization dynamics using the transverse magneto-optic Kerr effect in the energy range of 30-72 eV
-
- Jana, Somnath (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
- Terschlüsen, Joachim A. (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
- Stefanuik, Robert (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
visa fler...
-
- Plogmaker, Stefan (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik,MAX VI Lab, Lund, Sweden.
-
- Troisi, Sara (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
- Malik, Rameez S. (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
- Svanqvist, Mathias (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik,Swedish Def Res Agcy, FOI, SE-14725 Tumba, Sweden.
-
- Knut, Ronny (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
- Söderström, Johan, 1978- (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
- Karis, Olof (författare)
- Uppsala universitet,Molekyl- och kondenserade materiens fysik
-
visa färre...
-
(creator_code:org_t)
- AMER INST PHYSICS, 2017
- 2017
- Engelska.
-
Ingår i: Review of Scientific Instruments. - : AMER INST PHYSICS. - 0034-6748 .- 1089-7623. ; 88:3
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- In this paper, we present a spectrometer that is designed for element-specific and time-resolved transverse magneto-optic Kerr effect experiments at the high-harmonic generation pump-probe facility High Energy Laser Induced Overtone Source (HELIOS) laboratory. HELIOS delivers photons with energies between 30 eV and 72 eV with an overall time resolution of less than 40 fs. The spectrometer is based on a Rowland-circle geometry and allows for simultaneous measurements of all magnetic transition-metal elements. The setup also features easy sample transfer and alignment, and it combines high photon throughput, optimized data acquisition, and a fast switching of the magnetic field at the sample. The spectrometer performance is demonstrated by measuring the ultrafast demagnetization of permalloy. Our data are, for all practical purposes, identical to what have been reported in the earlier high-order harmonic generation work of a similar sample by Mathias et al. [Proc. Natl. Acad. Sci. U.S.A. 109, 4792-4797 (2012)], however, obtained within 15% of the acquisition time compared to their study. Furthermore, our data show a shift of the demagnetization curve of Ni relative to Fe, which has previously been interpreted as a delay of the Ni demagnetization to that of Fe [S. Mathias et al., Proc. Natl. Acad. Sci. U. S. A. 109, 4792-4797 (2012)].
Ämnesord
- NATURVETENSKAP -- Fysik -- Acceleratorfysik och instrumentering (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Accelerator Physics and Instrumentation (hsv//eng)
- NATURVETENSKAP -- Fysik -- Den kondenserade materiens fysik (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Condensed Matter Physics (hsv//eng)
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas
- Av författaren/redakt...
-
Jana, Somnath
-
Terschlüsen, Joa ...
-
Stefanuik, Rober ...
-
Plogmaker, Stefa ...
-
Troisi, Sara
-
Malik, Rameez S.
-
visa fler...
-
Svanqvist, Mathi ...
-
Knut, Ronny
-
Söderström, Joha ...
-
Karis, Olof
-
visa färre...
- Om ämnet
-
- NATURVETENSKAP
-
NATURVETENSKAP
-
och Fysik
-
och Acceleratorfysik ...
-
- NATURVETENSKAP
-
NATURVETENSKAP
-
och Fysik
-
och Den kondenserade ...
- Artiklar i publikationen
-
Review of Scient ...
- Av lärosätet
-
Uppsala universitet