Sökning: onr:"swepub:oai:gup.ub.gu.se/183953" >
Load applied on bon...
Load applied on bone-anchored transfemoral prosthesis: Characterization of a prosthesis-A pilot study
-
Frossard, L. (författare)
-
- Häggström, Eva (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
- Hagberg, Kerstin, 1957 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
visa fler...
-
- Brånemark, Rickard, 1960 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för ortopedi,Institute of Clinical Sciences, Department of Orthopaedics
-
visa färre...
-
(creator_code:org_t)
- Journal of Rehabilitation Research & Development, 2013
- 2013
- Engelska.
-
Ingår i: Journal of Rehabilitation Research and Development. - : Journal of Rehabilitation Research & Development. - 0748-7711. ; 50:5, s. 619-634
- Relaterad länk:
-
https://gup.ub.gu.se... (primary) (free)
-
visa fler...
-
https://doi.org/10.1...
-
https://gup.ub.gu.se...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- The objectives of this study were to (1) record the inner-prosthesis loading during activities of daily living (ADLs), (2) present a set of variables comparing loading data, and (3) provide an example of characterization of two prostheses. The load was measured at 200 Hz using a multi-axial transducer mounted between the residuum and the knee of an individual with unilateral transfemoral amputation fitted with a bone-anchored prosthesis. The load was measured while using two different prosthetic knees, mechanical (PRO1) and microprocessor-controlled (PRO2), during six ADLs. The characterization of the prostheses was achieved using a set of variables split into four categories, including temporal characteristics, maximum loading, loading slopes, and impulse. Approximately 360 gait cycles were analyzed for each prosthesis. PRO I showed a cadence improved by 19% and 7%, a maximum force on the long axis reduced by 11% and 19%, and an impulse reduced by 32% and 15% during descent of incline and stairs compared with PRO2, respectively. This work confirmed that the proposed apparatus and characterization can reveal how changes of prosthetic components are translated into inner-prosthetic loading.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Hälsovetenskap -- Sjukgymnastik (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Health Sciences -- Physiotherapy (hsv//eng)
Nyckelord
- activity of daily living
- artificial limb
- bone-anchorage
- gait
- impulse
- loading
- osseointegration
- TRANS-FEMORAL AMPUTEE
- KNEE-JOINTS
- OSSEOINTEGRATED FIXATION
- AMPUTATION
- PROSTHESES
- GAIT ANALYSIS
- DAILY-LIFE
- WALKING
- LIMB
- REHABILITATION
- PERFORMANCE
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas