Potential for lower absorbed dose in digital mammography: A JAFROC experiment using clinical hybrid images with simulated dose reduction

Timberg, Pontus (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Ruschin, Mark (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa fler...
Hemdal, Bengt (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Andersson, Ingvar (författare)
Lund University,Lunds universitet,Institutionen för translationell medicin,Medicinska fakulteten,Department of Translational Medicine,Faculty of Medicine
Mattsson, Sören (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Chakraborty, D. (författare)
Saunders, R. (författare)
Samei, E. (författare)
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
SPIE, 2006
2006
Engelska.
Ingår i: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. - : SPIE. - 1605-7422. ; 6146, s. 14614-14614
  • Tidskriftsartikel (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • To determine how image quality linked to tumor detection is affected by reducing the absorbed dose to 50% and 30% of the clinical levels represented by an average glandular dose (AGO) level of 1.3 mGy for a standard breast according to European guidelines. Materials and methods: 90 normal, unprocessed images were acquired from the screening department using a full-field digital mammography (FFDM) unit Mammomat Novation (Siemens). Into 40 of these, one to three simulated tumors were inserted per image at various positions. These tumors represented irregular-shaped malignant masses. Dose reduction was simulated in all 90 images by adding simulated quantum noise to represent images acquired at 50% and 30% of the original dose, resulting in 270 images, which were subsequently processed for final display. Four radiologists participated in a free-response receiver operating characteristics (FROG) study in which they searched for and marked suspicious positions of the masses as well as rated their degree of suspicion of occurrence on a one to four scale. Using the jackknife FROG (JAFROC) method, a score between 0 and 1 (where 1 represents best performance), referred to as a figure-of-merit (FOM), was calculated for each dose level. Results: The FOM was 0.73, 0.70, and 0.68 for the 100%, 50% and 30% dose levels, respectively. Using Analysis of the Variance (ANOVA) to test for statistically significant differences between any two of the three FOMs revealed that they were not statistically distinguishable (p-value of 0.26). Conclusion: For the masses used in this experiment, there was no significant change in detection by increasing quantum noise, thus indicating a potential for dose reduction.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Average glandular dose
Digital mammography
Dose reduction
FROC
Hybrid images
JAFROC
average glandular dose
digital mammography
dose reduction
hybrid images

Publikations- och innehållstyp

vet (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy