Sökning: onr:"swepub:oai:gup.ub.gu.se/235511" >
The effect of fixed...
The effect of fixed eye adaptation when using displays with a high luminance range
-
- Sund, Patrik (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
- Månsson, Lars Gunnar, 1951 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
- Båth, Magnus, 1974 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
(creator_code:org_t)
- SPIE, 2012
- 2012
- Engelska.
-
Ingår i: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. - : SPIE. - 1605-7422. ; 8318
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Calibration of medical review displays according to the part 14 Grayscale Standard Display Function (GSDF) is important in order to obtain consistency in displayed image quality since display technology and viewing conditions may vary substantially. Unfortunately, the purpose of the GSDF calibration is best suited for low luminance range conditions but is not optimal when using modern displays with a high luminance range. Low contrast objects will then obtain a greater visibility in mid-gray areas compared to similar objects in bright or dark regions. In this study, low contrast sinusoidal patterns were displayed on a high luminance range monitor under realistic viewing conditions. In order to simulate the viewing of an x-ray image with both dark and bright regions displayed simultaneously, the luminance of the patterns ranged from 2 to 600 cd/m2 while the observers were always adapted to the logarithmic average of 35 cd/m2. The results show a clear relationship between the patterns deviation from the adaptation luminance level and the necessary contrast required to detect the pattern. The results also indicate the potential for an improvement in the lowcontrast detectability over a large luminance range by adjusting the GSDF for the limited eye adaptation.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Image Processing (hsv//eng)
Publikations- och innehållstyp
- vet (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas