Comparison of two digital radiostereometric analysis methods in the determination of femoral head penetration in a total hip replacement phantom
-
- Bragdon, Charles R, 1959 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för ortopedi,Institute of Surgical Sciences, Department of Orthopaedics
-
Estok, D. M. (författare)
-
- Malchau, Henrik, 1951 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för ortopedi,Institute of Surgical Sciences, Department of Orthopaedics
-
visa fler...
-
- Kärrholm, Johan, 1951 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för de kirurgiska disciplinerna, Avdelningen för ortopedi,Institute of Surgical Sciences, Department of Orthopaedics
-
Yuan, X. (författare)
-
Bourne, R. (författare)
-
Veldhoven, J. (författare)
-
Harris, W. H. (författare)
-
visa färre...
-
(creator_code:org_t)
- 2006
- 2004
- Engelska.
-
Ingår i: J Orthop Res. - : Wiley. - 0736-0266. ; 22:3, s. 659-64
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Radiostereometric analysis (RSA) has been used extensively to evaluate the magnitude and direction of penetration of the femoral head into the acetabular component of a total hip replacement as a result of polyethylene wear and creep. The accuracy and precision of an RSA study depends on several factors, including the radiographic technique, the analytical software, and the positioning of the tantalum markers. This study had three sequential purposes. First, an in vitro phantom model was used to quantify the accuracy and precision of digital images versus conventional radiography in RSA measurements of penetration of the femoral head into the acetabular shell in a total hip replacement. The Umea RSA software package was used for analysis of both the conventional films, which were digitized at a resolution of 300 DPI, and digital radiographs, which were converted from a DICOM format at a resolution of 218 DPI. Digital radiography was found to be superior. Next, two methods of RSA analysis currently in use for determining femoral head penetration into polyethylene of total joint replacements were compared. Using the phantom model, we compared the Umea RSA system (Biomedical Innovations AB) to the RSA-CMS (RSA Clinical Measurement Solution) and in both cases used the digital radiographs. The Umea RSA system was found to be superior. Finally, two methods of marking the position of the acetabular component with tantalum beads were compared: one in which beads were inserted into previously described towers protruding from the back of the acetabular shell and another in which beads were placed into the peripheral flange of the polyethylene liner using the Umea RSA analysis system of the digital radiographs. The results using the two marker configurations were similar.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Kirurgi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Surgery (hsv//eng)
Nyckelord
- *Arthroplasty
- Replacement
- Hip
- Femur Head/*radiography
- Humans
- Phantoms
- Imaging
- *Photogrammetry
- Polyethylene
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas
- Av författaren/redakt...
-
Bragdon, Charles ...
-
Estok, D. M.
-
Malchau, Henrik, ...
-
Kärrholm, Johan, ...
-
Yuan, X.
-
Bourne, R.
-
visa fler...
-
Veldhoven, J.
-
Harris, W. H.
-
visa färre...
- Om ämnet
-
- MEDICIN OCH HÄLSOVETENSKAP
-
MEDICIN OCH HÄLS ...
-
och Klinisk medicin
-
och Kirurgi
- Artiklar i publikationen
-
J Orthop Res
-
Journal of Ortho ...
- Av lärosätet
-
Göteborgs universitet