Sökning: onr:"swepub:oai:lup.lub.lu.se:148999f6-eef5-4403-b171-b7f65e77d222" >
Continuous mechanic...
Continuous mechanical chest compression during in-hospital cardiopulmonary resuscitation of patients with pulseless electrical activity
-
Bonnemeier, Hendrik (författare)
-
Simonis, Gregor (författare)
-
- Olivecrona, Göran (författare)
- Lund University,Lunds universitet,Kardiologi,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Cardiology,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa fler...
-
Weidtmann, Britta (författare)
-
- Götberg, Matthias (författare)
- Lund University,Lunds universitet,Kardiologi,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Cardiology,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
-
Weitz, Gunther (författare)
-
Gerling, Ivana (författare)
-
Strasser, Ruth (författare)
-
Frey, Norbert (författare)
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2011
- 2011
- Engelska.
-
Ingår i: Resuscitation. - : Elsevier BV. - 1873-1570 .- 0300-9572. ; 82:2, s. 155-159
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Survival after in-hospital pulseless electrical activity (PEA) cardiac arrest is poor and has not changed during the last 10 years. Effective chest compressions may improve survival after PEA. We investigated whether a mechanical device (LUCAS (TM)-CPR) can ensure chest compressions during cardiac arrest according to guidelines and without interruption during transport, diagnostic procedures and in the catheter laboratory. Methods: We studied mechanical chest compression in 28 patients with PEA (pulmonary embolism (PE) n = 14; cardiogenic shock/acute myocardial infarction; n=9; severe hyperkalemia; n=2; sustained ventricular arrhythmias/electrical storm; n=3) in a university hospital setting. Results: During or immediately after CPR, 21 patients underwent coronary angiography and or pulmonary angiography. Successful return of a spontaneous circulation (ROSC) was achieved in 27 out of the 28 patients. Ten patients died within the first hour and three patients died within 2411 after CPR. A total of 14 patients survived and were discharged from hospital (13 without significant neurological deficit). Interestingly, six patients with PE did not have thrombolytic therapy due to contraindications. CT-angiography findings in these patients showed fragmentation of the thrombus suggesting thrombus breakdown as an additional effect of mechanical chest compressions. No patients exhibited any life-threatening device-related complications. Conclusion: Continuous chest compression with an automatic mechanical device is feasible, safe, and might improve outcomes after in-hospital-resuscitation of PEA. Patients with PE may benefit from effective continuous chest compression, probably due to thrombus fragmentation and increased pulmonary artery blood flow. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Anestesi och intensivvård (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Anesthesiology and Intensive Care (hsv//eng)