Sökning: onr:"swepub:oai:lup.lub.lu.se:19cbda62-1bba-482b-a0f4-7b9e661385c6" >
MR evaluation ex vi...
MR evaluation ex vivo and in vivo of a covered stent-graft for abdominal aortic aneurysms: ferromagnetism, heating, artifacts, and velocity mapping
-
Engellau, Lena (författare)
-
- Olsrud, Johan (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Brockstedt, Sara (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa fler...
-
- Albrechtsson, Ulf (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Norgren, Lars (författare)
- Lund University,Lunds universitet,Institutionen för kliniska vetenskaper, Malmö,Medicinska fakulteten,Department of Clinical Sciences, Malmö,Faculty of Medicine
-
- Ståhlberg, Freddy (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
-
- Larsson, Elna-Marie (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- 2000
- 2000
- Engelska.
-
Ingår i: Journal of Magnetic Resonance Imaging. - 1522-2586. ; 12:1, s. 112-121
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Magnetic resonance imaging (MRI) safety was evaluated at 1.5 T in a covered nickel titanium stent-graft (Vanguard) used for endovascular treatment of abdominal aortic aneurysms (AAAs). Imaging artifacts were assessed on MRI with contrast-enhanced (CE) three-dimensional (3D) MR angiography (MRA) and spiral computed tomography (CT) in 10 patients as well as ex vivo. Velocity mapping was performed in the suprarenal aorta and femoral arteries in 14 patients before and after stent-graft placement. For comparison it was also performed in six healthy volunteers. No ferromagnetism or heating was detected. Metal artifacts caused minimal image distortion on MRI/MRA. The artifacts disturbed image evaluation on CT at the graft bifurcation and graft limb junction. No significant differences in mean flow were found in patients before and after stent-graft placement. Our study indicates that MRI at 1.5 T may be performed safely in patients with the (Vanguard) stent-graft. MRI/MRA provides diagnostic image information. Velocity mapping is not included in our routine protocol.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- aneurysm
- aortic
- aorta
- grafts and prostheses
- MR imaging
- safety
- artifacts
- blood
- flow dynamics
Publikations- och innehållstyp