A comparative study of virtual hand prosthesis control using an inductive tongue control system
-
- Johansen, Daniel (författare)
- Aalborg University
-
- Sebelius, Fredrik (författare)
- Lund University,Lunds universitet,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
-
- Jensen, Stig (författare)
- Sahva A/S
-
visa fler...
-
- Bentsen, Bo (författare)
- Aalborg University
-
- Popović, Dejan B. (författare)
- University of Belgrade
-
- Andreasen Struijk, Lotte N S (författare)
- Aalborg University
-
visa färre...
-
(creator_code:org_t)
- 2015-10-19
- 2016
- Engelska 8 s.
-
Ingår i: Assistive Technology. - : Informa UK Limited. - 1040-0435 .- 1949-3614. ; 28:1, s. 22-29
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- This study compares the time required to activate a grasp or function of a hand prosthesis when using an electromyogram (EMG) based control scheme and when using a control scheme combining EMG and control signals from an inductive tongue control system (ITCS). Using a cross-over study design, 10 able-bodied subjects used a computer model of a hand and completed simulated grasping exercises. The time required to activate grasps was recorded and analyzed for both control schemes. End session mean activation times (ATs; seconds) for the EMG control scheme grasps 1 -5 were 0.80, 1.51, 1.95, 2.93, and 3.42; for the ITCS control scheme grasps 1 5 they were 1.19, 1.89, 1.75, 2.26, and 1.80. Mean AT for grasps 1 and 2 was statistically significant in favor of the EMG control scheme (p = 0.030; p = 0.004). For grasp 3 no statistical significance occurred, and for grasps 4 and 5 there was a statistical significance in favour of the ITCS control scheme (p = 0.048; p = 0.004). Based on the amount of training and the achieved level of performance, it is concluded that the proposed ITCS control scheme can be used as a means of enhancing prosthesis control.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)
Nyckelord
- Myoelectric control
- Prosthesis control schemes
- Upper limb prosthetics
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas