Prediction of stopping-power ratios in flattening-filter free beams.
-
- Ceberg, Crister (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
Johnsson, Stefan (författare)
-
- Dalaryd, Mårten (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa fler...
-
- Knöös, Tommy (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- 2010-02-18
- 2010
- Engelska.
-
Ingår i: Medical Physics. - : Wiley. - 0094-2405. ; 37:3, s. 1164-1168
- Relaterad länk:
-
https://portal.resea... (primary) (free)
-
visa fler...
-
http://www.ncbi.nlm....
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- PURPOSE: In recent years, there has been an increasing interest in flattening-filter free (FFF) beams. However, since the removal of the flattening filter will affect both the mean and the variance of the energy spectrum, current beam-quality specifiers may not be adequate for reference dosimetry in such beams. The purpose of this work was to investigate an alternative, more general beam-quality specifier. METHODS: The beam-quality specifier used in this work was a combination of the kerma-weighted mean and the coefficient of variation of the linear attenuation coefficient in water. These parameters can in theory be determined from narrow-beam transmission measurements using a miniphantom "in-air," which is a measurement condition well suited also to small and nonstandard fields. The relation between the Spencer-Attix stopping-power ratios and this novel beam-quality specifier was described by a simple polynomial. For reference, the authors used Monte Carlo calculated spectra and stopping-power data for nine different beams, with and without flattening filter. RESULTS: The polynomial coefficients were obtained by least-squares optimization. For all beams included in this investigation, the average of the differences between the predicted and the Monte Carlo calculated stopping-power ratios was 0.02 +/- 0.17% (1 SD) (including TomoTherapy and CyberKnife example beams). CONCLUSIONS: An alternative dual-parameter beam-quality specifier was investigated. The evaluation suggests that it can be used successfully to predict stopping-power ratios in FFF as well as conventional beams, regardless of filtration.
Ämnesord
- NATURVETENSKAP -- Biologi -- Biofysik (hsv//swe)
- NATURAL SCIENCES -- Biological Sciences -- Biophysics (hsv//eng)
Nyckelord
- dosimetry
- phantoms
- Monte Carlo methods
- radiation therapy