Combination of a leadless pacemaker and subcutaneous defibrillator with nine effective shock treatments during follow-up of 18 months
-
- Ljungström, Erik (författare)
- Lund University,Skåne University Hospital
-
- Brandt, Johan (författare)
- Lund University,Lunds universitet,Thoraxkirurgi,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Thoracic Surgery,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine,Skåne University Hospital
-
- Mörtsell, David (författare)
- Lund University,Lunds universitet,Arrhytmias and Cardiac Device treatment,Forskargrupper vid Lunds universitet,Lund University Research Groups,Skåne University Hospital
-
visa fler...
-
- Borgquist, Rasmus (författare)
- Lund University,Lunds universitet,Arrhytmias and Cardiac Device treatment,Forskargrupper vid Lunds universitet,Lund University Research Groups,Skåne University Hospital
-
- Wang, Lingwei (författare)
- Lund University,Lunds universitet,Arrhytmias and Cardiac Device treatment,Forskargrupper vid Lunds universitet,Lund University Research Groups,Skåne University Hospital
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2019
- 2019
- Engelska 3 s.
-
Ingår i: Journal of Electrocardiology. - : Elsevier BV. - 0022-0736. ; 56, s. 1-3
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- We present a case of combination of a leadless pacemaker (Micra) and a subcutaneous implantable cardioverter-defibrillator (S-ICD). The patient had a total of nine adequate shock treatments of ventricular fibrillation during 18 months of follow-up after the implantation. The shock treatments did not lead to any alteration in the Micra. All three sensing vectors of the S-ICD worked well. After 18 months, the functioning of both Micra and S-ICD continues to be uneventful. This case demonstrates that S-ICD combined with Micra may be a safe and feasible approach to provide pacing and ICD treatment without intracardiac leads.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Kardiologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk apparatteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Equipment Engineering (hsv//eng)
Nyckelord
- Effective shock treatment
- Leadless pacemaker
- Subcutaneous defibrillator (S-ICD)
- Ventricular fibrillation (VF)