Manual correction of semi-automatic three-dimensional echocardiography is needed for right ventricular assessment in adults; validation with cardiac magnetic resonance
-
- Ostenfeld, Ellen (författare)
- Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Kardiologiska klinikens forskargrupp,Forskargrupper vid Lunds universitet,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Cardiology Research Group,Lund University Research Groups
-
- Carlsson, Marcus (författare)
- Lund University,Lunds universitet,Klinisk fysiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Physiology (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
Shahgaldi, Kambiz (författare)
-
visa fler...
-
- Roijer, Anders (författare)
- Lund University,Lunds universitet,Kardiologi,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Cardiology,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Holm, Johan (författare)
- Lund University,Lunds universitet,Kardiologi,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Cardiology,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- 2012-01-06
- 2012
- Engelska.
-
Ingår i: Cardiovascular Ultrasound. - : Springer Science and Business Media LLC. - 1476-7120. ; 10
- Relaterad länk:
-
https://portal.resea... (primary) (free)
-
visa fler...
-
http://dx.doi.org/10... (free)
-
https://cardiovascul...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Background: Three-dimensional echocardiography (3DE) and semi-automatic right ventricular delineation has been proposed as an appropriate method for right ventricle (RV) evaluation. We aimed to examine how manual correction of semi-automatic delineation influences the accuracy of 3DE for RV volumes and function in a clinical adult setting using cardiac magnetic resonance (CMR) as the reference method. We also examined the feasibility of RV visualization with 3DE. Methods: 62 non-selected patients were examined with 3DE (Sonos 7500 and iE33) and with CMR (1.5T). Endocardial RV contours of 3DE-images were semi-automatically assessed and manually corrected in all patients. End-diastolic (EDV), end-systolic (ESV) volumes, stroke volume (SV) and ejection fraction (EF) were computed. Results: 53 patients (85%) had 3DE-images feasible for examination. Correlation coefficients and Bland Altman biases between 3DE with manual correction and CMR were r = 0.78, -22 +/- 27 mL for EDV, r = 0.83, -7 +/- 16 mL for ESV, r = 0.60, -12 +/- 18 mL for SV and r = 0.60, -2 +/- 8% for EF (p < 0.001 for all r-values). Without manual correction r-values were 0.77, 0.77, 0.70 and 0.49 for EDV, ESV, SV and EF, respectively (p < 0.001 for all r-values) and biases were larger for EDV, SV and EF (-32 +/- 26 mL, -21 +/- 15 mL and -6 +/- 9%, p <= 0.01 for all) compared to manual correction. Conclusion: Manual correction of the 3DE semi-automatic RV delineation decreases the bias and is needed for acceptable clinical accuracy. 3DE is highly feasible for visualizing the RV in an adult clinical setting.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Kardiologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)
Nyckelord
- volumes
- function
- right ventricle
- magnetic resonance
- three-dimensional echocardiography
- clinical
- adult
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas