Comparison of conventional and Si-photomultiplier-based PET systems for image quality and diagnostic performance
-
- Oddstig, Jenny (författare)
- Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Nuclear medicine, Malmö,Lund University Research Groups,Skåne University Hospital
-
- Leide Svegborn, Sigrid (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Skåne University Hospital
-
- Almquist, Helen (författare)
- Skåne University Hospital
-
visa fler...
-
- Bitzén, Ulrika (författare)
- Skåne University Hospital
-
- Garpered, Sabine (författare)
- Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,Skåne University Hospital
-
- Hedeer, Fredrik (författare)
- Skåne University Hospital
-
- Hindorf, Cecilia (författare)
- Lund University,Lunds universitet,Hjärt-MR-gruppen i Lund,Forskargrupper vid Lunds universitet,Lund Cardiac MR Group,Lund University Research Groups,Skåne University Hospital
-
- Jögi, Jonas (författare)
- Lund University,Lunds universitet,Hjärt-MR-gruppen i Lund,Forskargrupper vid Lunds universitet,Lund Cardiac MR Group,Lund University Research Groups,Skåne University Hospital
-
- Jönsson, Lena (författare)
- Lund University,Lunds universitet,Nuclear Medicine Physics,Forskargrupper vid Lunds universitet,Lund University Research Groups,Skåne University Hospital
-
- Minarik, David (författare)
- Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Nuclear medicine, Malmö,Lund University Research Groups,Skåne University Hospital
-
- Petersson, Richard (författare)
- Skåne University Hospital
-
- Welinder, Annika (författare)
- Skåne University Hospital
-
- Wollmer, Per (författare)
- Lund University,Lunds universitet,Klinisk fysiologi och nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,Clinical Physiology and Nuclear Medicine, Malmö,Lund University Research Groups,Skåne University Hospital
-
- Trägårdh, Elin (författare)
- Lund University,Lunds universitet,Nuklearmedicin, Malmö,Forskargrupper vid Lunds universitet,WCMM- Wallenberg center för molekylär medicinsk forskning,Medicinska fakulteten,Nuclear medicine, Malmö,Lund University Research Groups,WCMM-Wallenberg Centre for Molecular Medicine,Faculty of Medicine,Skåne University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2019-10-22
- 2019
- Engelska.
-
Ingår i: BMC Medical Imaging. - : Springer Science and Business Media LLC. - 1471-2342. ; 19:1
- Relaterad länk:
-
http://dx.doi.org/10... (free)
-
visa fler...
-
https://bmcmedimagin...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- BACKGROUND: A new generation of positron emission tomography with computed tomography (PET-CT) was recently introduced using silicon (Si) photomultiplier (PM)-based technology. Our aim was to compare the image quality and diagnostic performance of a SiPM-based PET-CT (Discovery MI; GE Healthcare, Milwaukee, WI, USA) with a time-of-flight PET-CT scanner with a conventional PM detector (Gemini TF; Philips Healthcare, Cleveland, OH, USA), including reconstruction algorithms per vendor's recommendations. METHODS: Imaging of the National Electrical Manufacturers Association IEC body phantom and 16 patients was carried out using 1.5 min/bed for the Discovery MI PET-CT and 2 min/bed for the Gemini TF PET-CT. Images were analysed for recovery coefficients for the phantom, signal-to-noise ratio in the liver, standardized uptake values (SUV) in lesions, number of lesions and metabolic TNM classifications in patients. RESULTS: In phantom, the correct (> 90%) activity level was measured for spheres ≥17 mm for Discovery MI, whereas the Gemini TF reached a correct measured activity level for the 37-mm sphere. In patient studies, metabolic TNM classification was worse using images obtained from the Discovery MI compared those obtained from the Gemini TF in 4 of 15 patients. A trend toward more malignant, inflammatory and unclear lesions was found using images acquired with the Discovery MI compared with the Gemini TF, but this was not statistically significant. Lesion-to-blood-pool SUV ratios were significantly higher in images from the Discovery MI compared with the Gemini TF for lesions smaller than 1 cm (p < 0.001), but this was not the case for larger lesions (p = 0.053). The signal-to-noise ratio in the liver was similar between platforms (p = 0.52). Also, shorter acquisition times were possible using the Discovery MI, with preserved signal-to-noise ratio in the liver. CONCLUSIONS: Image quality was better with Discovery MI compared to conventional Gemini TF. Although no gold standard was available, the results indicate that the new PET-CT generation will provide potentially better diagnostic performance.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
- NATURVETENSKAP -- Fysik -- Annan fysik (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Other Physics Topics (hsv//eng)
Nyckelord
- Digital PET
- FDG
- Image quality
- Oncology
- PET-CT
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas
- Av författaren/redakt...
-
Oddstig, Jenny
-
Leide Svegborn, ...
-
Almquist, Helen
-
Bitzén, Ulrika
-
Garpered, Sabine
-
Hedeer, Fredrik
-
visa fler...
-
Hindorf, Cecilia
-
Jögi, Jonas
-
Jönsson, Lena
-
Minarik, David
-
Petersson, Richa ...
-
Welinder, Annika
-
Wollmer, Per
-
Trägårdh, Elin
-
visa färre...
- Om ämnet
-
- MEDICIN OCH HÄLSOVETENSKAP
-
MEDICIN OCH HÄLS ...
-
och Klinisk medicin
-
och Radiologi och bi ...
-
- NATURVETENSKAP
-
NATURVETENSKAP
-
och Fysik
-
och Annan fysik
- Artiklar i publikationen
-
BMC Medical Imag ...
- Av lärosätet
-
Lunds universitet