THE EFFECT OF REDUCED BREAST COMPRESSION IN BREAST TOMOSYNTHESIS: HUMAN OBSERVER STUDY USING CLINICAL CASES.

Förnvik, Daniel (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
Andersson, Ingvar (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Svahn, Tony (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa fler...
Timberg, Pontus (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Medicinsk strålningsfysik, Malmö,Radiology Diagnostics, Malmö,Lund University Research Groups,Medical Radiation Physics, Malmö
Zackrisson, Sophia (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Tingberg, Anders (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
2010-03-11
2010
Engelska.
Ingår i: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; Apr 7, s. 118-123
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The aim of this study was to investigate whether the compression force used with conventional mammography can be reduced with breast tomosynthesis (BT), without adversely affecting the visualisation of normal and pathological structures. Forty-five women were examined with BT using full (same as for 2D mammography) and half compression force. Both examinations were performed with the same acquisition parameters. A total of 103 paired structure images were evaluated according to specified image quality criteria. Three experienced radiologists participated in the study. They had to make a forced choice, i.e. choose the image they felt best fulfilled the image quality criteria. The results showed no evident difference in the image quality, indicating that BT may be performed with substantially less compression force compared with 2D mammography. A majority of the examined women felt that half compression was more comfortable than full compression.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy