Assay discrepancy in mild haemophilia A

Armstrong, Elina (författare)
Helsinki University Central Hospital
Hillarp, Andreas (författare)
Lund University,Lunds universitet,Klinisk kemi, Malmö,Forskargrupper vid Lunds universitet,Clinical Chemistry, Malmö,Lund University Research Groups,Regional Laboratories Region Skåne
 (creator_code:org_t)
2014-06-24
2014
Engelska.
Ingår i: European Journal of Haematology. Supplementum. - : Wiley. - 0902-4506 .- 0902-4441. ; 93:s76, s. 48-50
  • Forskningsöversikt (refereegranskat)
Abstract Ämnesord
Stäng  
  • UNLABELLED: There are three main methods used to assay factor VIII (FVIII) activity: the one-stage and two-stage clotting assays and the two-stage chromogenic method. The most commonly used assay for the diagnosis of haemophilia A is the automated one-stage FVIII assay. The classical two-stage FVIII assays are less frequently used. The chromogenic FVIII:C assay is a variant of the two-stage assay. It is easier to use and therefore used more commonly. Recently significant assay discrepancy has been recognised in the FVIII:C measurements in approximately one-third of mild haemophilia A patients. This so-called discrepant mild haemophilia A is characterised by a high ratio of one-stage/two-stage assay with one-stage FVIII levels that are typically more than double those of the two-stage coagulation assay. There are several mutations that destabilise the FVIIIa structure that can explain this result of a more pronounced decrease of the chromogenic FVIII:C activity compared with the one-stage activity. These mutations are clustered at the interfaces of the A1, A2 and A3 domains of the FVIII protein. The inverse discrepancy, where the one-stage assay gives lower FVIII:C results than the chromogenic assay, seems to be associated with mutations found close to important sites for thrombin cleavage or FIX binding. We are carrying out a study of mild haemophilia A samples from the Malmö Haemophilia Centre of families with a unique F8 genotype. The activity of FVIII will be measured using a chromogenic assay and two different one-stage assays. We hope to estimate the true size of assay discrepancy.AIM: This project will review assay discrepancy in mild/moderate haemophilia A and the risk of misdiagnosis. The overall aim is to estimate the size of the problem and to learn from the literature and experiences from our centre as well as to suggest recommendations on how to avoid misdiagnosis.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Hematologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Hematology (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Medicinska och farmaceutiska grundvetenskaper -- Läkemedelskemi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Basic Medicine -- Medicinal Chemistry (hsv//eng)

Nyckelord

Blood Coagulation Tests/methods
Hemophilia A/blood
Humans
Reproducibility of Results
Severity of Illness Index

Publikations- och innehållstyp

for (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Armstrong, Elina
Hillarp, Andreas
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Hematologi
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Medicinska och f ...
och Läkemedelskemi
Artiklar i publikationen
European Journal ...
Av lärosätet
Lunds universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy