Charcoal as an airway isoflurane reflection filter
-
Dahm, S L (författare)
-
Steptoe, P (författare)
-
- Luttropp, Hans-Henrik (författare)
- Lund University,Lunds universitet,Anestesiologi och intensivvård,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Anesthesiology and Intensive Care,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa fler...
-
- Reinstrup, Peter (författare)
- Lund University,Lunds universitet,Anestesiologi och intensivvård,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Anesthesiology and Intensive Care,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- Ovid Technologies (Wolters Kluwer Health), 1998
- 1998
- Engelska.
-
Ingår i: European Journal of Anaesthesiology. - : Ovid Technologies (Wolters Kluwer Health). - 1365-2346 .- 0265-0215. ; 15:2, s. 230-233
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- The isoflurane-saving and CO2-retaining effects of a charcoal filter were compared with a Siemens standard heat and moisture (HME) exchanger and an emptied specimen (dummy). Isoflurane was delivered during the inspiratory phase and consumption investigated at 10, 15 and 25 cycles min-1. The investigation was performed by ventilation with humidified air with a constant end-tidal CO2 and temperature. For a comparison, isoflurane was delivered in a conventional manner via the ventilator. The arrangement with a charcoal filter reduced the isoflurane consumption by a factor of 2.0-2.6, depending on ventilatory rate. Most of the saving was a result of the method of isoflurane delivery (factor 1.4-2.0), while adding the reflector gave a further reduction (factor 1.3-1.5). One circumstance that reduced the net efficiency of the charcoal filter was that it also reflected CO2; consequently, total minute ventilation had to be increased to maintain constant end-tidal CO2.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Anestesi och intensivvård (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Anesthesiology and Intensive Care (hsv//eng)
Nyckelord
- anaesthetic
- volatile
- isoflurane
- equipment
- charcoal filter
- physics
- gas adsorption