Sökning: onr:"swepub:oai:lup.lub.lu.se:d8031fc5-b359-4012-804a-ff193b79f16c" >
Fabrication and Cha...
Fabrication and Characterization of Ultra-Thin PIN Silicon Detectors for Counting the Passage of MeV Ions
-
Abdel, Naseem (författare)
-
- Pallon, Jan (författare)
- Lund University,Lunds universitet,Kärnfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Nuclear physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
- Graczyk, Mariusz (författare)
- Lund University,Lunds universitet,Fasta tillståndets fysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Solid State Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
visa fler...
-
- Maximov, Ivan (författare)
- Lund University,Lunds universitet,Fasta tillståndets fysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Solid State Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH
-
- Wallman, Lars (författare)
- Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
-
visa färre...
-
(creator_code:org_t)
- 2013
- 2013
- Engelska.
-
Ingår i: IEEE Transactions on Nuclear Science. - 0018-9499. ; 60:2, s. 1182-1188
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- This paper describes the fabrication and initial characterization of an ultra-thin silicon PIN detector using a new technique in silicon nanotechnology. In collaboration with the Nuclear Physics Division and the Lund Nano Lab at Lund University, we have developed and manufactured ultra thin Delta E-detectors for spectroscopic applications. The fabrication process has been carried out using a double-polished silicon substrate n-type wafer and locally thinning by means of a 10:1 solution of 25% tetramethyl ammonium hydroxide (TMAH) with Isopropyl alcohol. More than 100 detectors of different thicknesses, down to 5 mu m with active areas ranging from 0.71 to 0.172 mm(2), have been fabricated. The main design considerations of our thin detectors were a very low leakage current below 12 nA and a low full depletion voltage at a reverse bias less than 1.5 V. Finally, most of our thin detectors offer an energy resolution (FWHM) as low as 31 keV for 5.487 MeV alpha particles from a Am-241 source.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Nanoteknik -- Nanoteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Nano-technology -- Nano-technology (hsv//eng)
- NATURVETENSKAP -- Fysik -- Acceleratorfysik och instrumentering (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Accelerator Physics and Instrumentation (hsv//eng)
Nyckelord
- Energy resolution
- leakage current
- silicon nanotechnology
- TMAH
- etching
- ultra-thin PIN detector
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas