Serum cystatin C, determined by a rapid, automated particle-enhanced turbidimetric method, is a better marker than serum creatinine for glomerular filtration rate
-
Kyhse-Andersen, Jan (författare)
-
Schmidt, Camilla (författare)
-
Nordin, Gunnar (författare)
-
visa fler...
-
Andersson, Britta (författare)
-
- Nilsson-Ehle, Peter (författare)
- Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
-
- Lindström, Veronica (författare)
- Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
-
- Grubb, Anders (författare)
- Lund University,Lunds universitet,Avdelningen för klinisk kemi och farmakologi,Institutionen för laboratoriemedicin,Medicinska fakulteten,Division of Clinical Chemistry and Pharmacology,Department of Laboratory Medicine,Faculty of Medicine
-
visa färre...
-
(creator_code:org_t)
- 1994
- 1994
- Engelska.
-
Ingår i: Clinical Chemistry. - 0009-9147. ; 40:10, s. 1921-1926
- Relaterad länk:
-
http://www.clinchem.... (free)
-
visa fler...
-
https://lup.lub.lu.s...
-
visa färre...
Abstract
Ämnesord
Stäng
- We describe a fully automated particle-enhanced turbidimetric assay for cystatin C in undiluted serum and EDTA-plasma. The throughput is 90 samples per hour and urgent samples can be analyzed in 7 min. The assay range (0.4-14.1 mg/L) covers the concentration range in health and disease. The within- and between-run imprecision is 0.9% and 2.2%, respectively. Analytical recovery of additions of recombinant cystatin C averaged 98%. Rheumatoid factors (< or = 323,000 IU/L), bilirubin (< or = 150 mumol/L), hemoglobin (< or = 1.2 g/L), and triglycerides (< or = 8.5 mmol/L) do not interfere in the assay. In view of the superior (by ROC analysis) diagnostic accuracy of serum concentrations of cystatin C for reduced glomerular filtration rate (GFR) in comparison with creatinine, cystatin C seems an attractive alternative to creatinine for estimation of GFR.