Bhogal, P (författare)
-
Brouwer, PA (författare)
-
- Yeo, L (författare)
- Karolinska Institutet
-
visa fler...
-
- Svensson, M (författare)
- Karolinska Institutet
-
- Soderman, M (författare)
- Karolinska Institutet
-
visa färre...
-
(creator_code:org_t)
- 2017-09-28
- 2018
- Engelska.
-
Ingår i: Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences. - : SAGE Publications. - 2385-2011. ; 24:1, s. 4-13
- Relaterad länk:
-
https://europepmc.or...
-
visa fler...
-
http://kipublication...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- The aim of this study was to report our single centre experience with the Medina Embolic Device (MED). Methods We performed a retrospective analysis of prospectively collected data to identify all patients treated with the MED. A total of 14 aneurysms (non-consecutive), in 13 patients, were treated including one ruptured and one partially thrombosed aneurysm. Fundus diameter was ≥5 mm in all cases. We evaluated the angiographic appearances, the clinical status, complications, and the need for adjunctive devices or repeat treatments. Results Aneurysm location was cavernous internal carotid artery (ICA; n = 1), supraclinoid ICA ( n = 1), terminal ICA ( n = 2), anterior communicating artery (AComA; n = 4), A2–3 ( n = 1), M1–2 junction ( n = 1), posterior communicating artery (PComA; n = 1), superior cerebellar artery (SCA; n = 1), and basilar tip ( n = 2). The average aneurysm fundus size was 8.6 mm (range 7–10 mm) and average neck size 3.75 mm (range 1.9–6.9 mm). Immediate angiographic results were modified Raymond–Roy occlusion classification (mRRC) I n = 2, mRRC II n = 1, mRRC IIIa n = 2, mRRC IIIb n = 2, the remaining 7 aneurysms showed complete opacification. At follow-up angiography (mean 5 months) mRRC I n = 5, mRRC II n = 5, mRRC IIIa n = 3, and persistent filling was seen in 1 aneurysm. Overall, four patients had repeat treatment and one is pending further treatment. Of the aneurysms treated with more than one MED, 75% showed complete occlusion at 6-month follow up whereas only one aneurysm treated with a single device showed complete occlusion. Overall, three patients had temporary complications and there were no deaths. Conclusions The MED is an intra-saccular flow-diverting device with satisfactory angiographic results and an acceptable safety profile. Use of a single MED cannot be recommended and further longer term studies are needed prior to widespread clinical use.
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas