Characterization of HPGe detectors using Computed Tomography
-
- Hedman, Angelica, 1984- (författare)
- Umeå universitet,Radiofysik,Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
-
- Gogani, J. B. (författare)
- Totalförsvarets forskningsinstitut (FOI),Swedish Defence Research Agency (FOI),Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
-
- Granstrom, M. (författare)
- Totalförsvarets forskningsinstitut (FOI),Swedish Defence Research Agency (FOI),Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
-
visa fler...
-
- Johansson, Lennart (författare)
- Umeå universitet,Radiofysik
-
- Andersson, Jonas S. (författare)
- Umeå universitet,Institutionen för strålningsvetenskaper
-
- Ramebäck, Henrik, 1967 (författare)
- Chalmers tekniska högskola,Chalmers University of Technology,Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden; Chalmers University of Technology, Department of Chemical and Biological Engineering, Nuclear Chemistry, SE-41296 Göteborg, Sweden
-
visa färre...
-
(creator_code:org_t)
- Elsevier BV, 2015
- 2015
- Engelska.
-
Ingår i: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. - : Elsevier BV. - 0168-9002 .- 1872-9576. ; 785, s. 21-25
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://research.cha...
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
visa färre...
Abstract
Ämnesord
Stäng
- Computed Tomography (CT) high resolution imaging have been used to investigate if there is a significant change in the crystal-to-window distance, i.e. the air gap thickness, in a small n-type detector cooled to 77 K, and in a medium sized p-type HPGe detector when cooled to 100 K. The findings were compared to detector dimension data made available by the manufacturer. The air gap thickness increased by (0.38 +/- 0.07) mm for the n-type detector and by (0.40 +/- 0.15) mm for the p-type detector when the detectors were cooled to 77 resp. 100 K compared to at room temperature. Monte Carlo calculations indicate that these differences have a significant impact on the efficiency in close geometries (
Ämnesord
- NATURVETENSKAP -- Fysik -- Subatomär fysik (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Subatomic Physics (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Image Processing (hsv//eng)
Nyckelord
- Semi-empirical calibration
- Monte Carlo modelling
- Gamma spectrometry
- Gespecor
- Computed Tomography
- HPGe