Characterization of HPGe detectors using Computed Tomography

Hedman, Angelica, 1984- (författare)
Umeå universitet,Radiofysik,Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
Gogani, J. B. (författare)
Totalförsvarets forskningsinstitut (FOI),Swedish Defence Research Agency (FOI),Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
Granstrom, M. (författare)
Totalförsvarets forskningsinstitut (FOI),Swedish Defence Research Agency (FOI),Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
visa fler...
Johansson, Lennart (författare)
Umeå universitet,Radiofysik
Andersson, Jonas S. (författare)
Umeå universitet,Institutionen för strålningsvetenskaper
Ramebäck, Henrik, 1967 (författare)
Chalmers tekniska högskola,Chalmers University of Technology,Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden; Chalmers University of Technology, Department of Chemical and Biological Engineering, Nuclear Chemistry, SE-41296 Göteborg, Sweden
visa färre...
 (creator_code:org_t)
Elsevier BV, 2015
2015
Engelska.
Ingår i: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. - : Elsevier BV. - 0168-9002 .- 1872-9576. ; 785, s. 21-25
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Computed Tomography (CT) high resolution imaging have been used to investigate if there is a significant change in the crystal-to-window distance, i.e. the air gap thickness, in a small n-type detector cooled to 77 K, and in a medium sized p-type HPGe detector when cooled to 100 K. The findings were compared to detector dimension data made available by the manufacturer. The air gap thickness increased by (0.38 +/- 0.07) mm for the n-type detector and by (0.40 +/- 0.15) mm for the p-type detector when the detectors were cooled to 77 resp. 100 K compared to at room temperature. Monte Carlo calculations indicate that these differences have a significant impact on the efficiency in close geometries (

Ämnesord

NATURVETENSKAP  -- Fysik -- Subatomär fysik (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences -- Subatomic Physics (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

Semi-empirical calibration
Monte Carlo modelling
Gamma spectrometry
Gespecor
Computed Tomography
HPGe

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy