- Hughes, Marion, et al.
(author)
-
Conformation of adjacent self-expanding stents: a cross-sectional in vitro study
- 2006
-
In: Cardiovascular and Interventional Radiology. - : Springer Science and Business Media LLC. - 1432-086X .- 0174-1551. ; 29:2, s. 255-259
-
Journal article (peer-reviewed)abstract
- We examined the proximal conformation of three commonly used self-expanding stents when the stents were deployed adjacent to one another in a tubular model, simulating a "kissing" stent technique. The stent pairs were evaluated by computed tomogrphy to determine the cross-sectional area excluded by the stents within the model. The mean areas associated with each stent pair were compared and significance evaluated by a t-test. A statistically significant difference was found when the area excluded by adjacent Wallstents was compared with both the Luminexx and SMART stents (p < 0.001 and p < 0.002, respectively). The difference in the area excluded and differences in conformation might play a role in the lower patencies that have been observed in "kissing" stent series.
|
|