SwePub
Sök i SwePub databas

  Extended search

Träfflista för sökning "WFRF:(Knöös Tommy) "

Search: WFRF:(Knöös Tommy)

  • Result 1-25 of 67
Sort/group result
   
EnumerationReferenceCoverFind
1.
  • Aarup, Lasse Rye, et al. (author)
  • The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverage
  • 2009
  • In: Radiotherapy and Oncology. - : Elsevier BV. - 1879-0887 .- 0167-8140. ; 91:3, s. 405-414
  • Journal article (peer-reviewed)abstract
    • Purpose: To evaluate against Monte-Carlo the performance of various dose calculations algorithms regarding lung turnout coverage in stereotactic body radiotherapy (SBRT) conditions. Materials and methods: Dose distributions in virtual lung phantoms have been calculated using four commercial Treatment Planning System (TPS) algorithms and one Monte Carlo (MC) system (EGSnrc). We compared the performance of the algorithms in calculating the target dose for different degrees of lung inflation. The phantoms had a cubic 'body' and 'lung' and a central 2-cm diameter spherical 'tumour' (the body and turnout have unit density). The lung tissue was assigned five densities (rho(lung)): 0.01, 0.1, 0.2, 0.4 and 1 g/cm(3). Four-field treatment plans were calculated with 6- and 18 MV narrow beams for each value of rho(lung). We considered the Pencil Beam Convolution (PBCEl) and the Analytical Anisotropic Algorithm (AAA(ECl)) from Varian Eclipse and the Pencil Beam Convolution (PBCOMP) and the Collapsed Cone Convolution (CCCOMP) algorithms from Oncentra MasterPlan. Results: When changing rho(lung) from 0.4 to 0.1 g/cm(3), the MC median target dose decreased from 89.2% to 74.9% for 6 MV and from 83.3% to 61.6% for 18 MV (of dose maximum in the homogenous case at both energies), while for both PB algorithms the median target dose was virtually independent of lung density. Conclusions: Both PB algorithms overestimated the target dose, the overestimation increasing as rho(lung) decreased. Concerning target dose, the AAA(ECl) and CCCOMP algorithms appear to be adequate alternatives to MC. (C) 2009 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and oncology 91 (2009) 405-414
  •  
2.
  • Ahnesjö, Anders, et al. (author)
  • Application of the convolution method for calculation of output factors for therapy photon beams
  • 1992
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 19:2, s. 295-301
  • Journal article (peer-reviewed)abstract
    • The output factor for a therapy photon beam is defined as the dose per monitor unit relative to the dose per monitor unit in a reference field. Convolution models for photon dose calculations yield the dose in units normalized to the incident energy fluence with phantom scatter intrinsically modeled. Output factors calculated with the convolution method as the dose per unit energy fluence relative to the calculated dose per unit energy fluence in a reference field could deviate as much as 5% if corrections are not made for perturbations due to treatment head scatter. Significant perturbations are particles backscattered from the collimators to the monitor and photons forward scattered from the filter and collimators in the treatment head. The forward scatter adds an "unmonitored" contribution to the total energy fluence of the beam. A model is developed that describes the field size dependence of these perturbations for conversion of output factors, calculated with the convolution method, to machine output factors as an integrated part in treatment planning. The necessary machine characteristics are derived from measurements of the output in air for a limited set of field sizes. The method has been tested using five different multileaf collimated irregular fields at 6 MV and for a large set of rectangular fields at 5, 6, and 18 MV and found to predict output factors with an accuracy better than 1%.
  •  
3.
  •  
4.
  • Björk, Peter, et al. (author)
  • Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy
  • 2000
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 27:11, s. 2580-2588
  • Journal article (peer-reviewed)abstract
    • The aim of the present study is to examine the validity of using silicon semiconductor detectors in degraded electron beams with a broad energy spectrum and a wide angular distribution. A comparison is made with diamond detector measurements, which is the dosimeter considered to give the best results provided that dose rate effects are corrected for. Two-dimensional relative absorbed dose distributions in electron beams (6-20 MeV) for intraoperative radiation therapy (IORT) are measured in a water phantom. To quantify deviations between the detectors, a dose comparison tool that simultaneously examines the dose difference and distance to agreement (DTA) is used to evaluate the results in low- and high-dose gradient regions, respectively. Uncertainties of the experimental measurement setup (+/- 1% and +/- 0.5 mm) are taken into account by calculating a composite distribution that fails this dose-difference and DTA acceptance limit. Thus, the resulting area of disagreement should be related to differences in detector performance. The dose distributions obtained with the diode are generally in very good agreement with diamond detector measurements. The buildup region and the dose falloff region show good agreement with increasing electron energy, while the region outside the radiation field close to the water surface shows an increased difference with energy. The small discrepancies in the composite distributions are due to several factors: (a) variation of the silicon-to-water collision stopping-power ratio with electron energy, (b) a more pronounced directional dependence for diodes than for diamonds, and (c) variation of the electron fluence perturbation correction factor with depth. For all investigated treatment cones and energies, the deviation is within dose-difference and DTA acceptance criteria of +/- 3% and +/- 1 mm, respectively. Therefore, p-type silicon diodes are well suited, in the sense that they give results in close agreement with diamond detectors, for practical measurements of relative absorbed dose distributions in degraded electron beams used for IORT.
  •  
5.
  • Björk, Peter, et al. (author)
  • Design and dosimetry characteristics of a soft-docking system for intraoperative radiation therapy
  • 2000
  • In: International Journal of Radiation Oncology, Biology, Physics. - 0360-3016. ; 47:2, s. 527-533
  • Journal article (peer-reviewed)abstract
    • PURPOSE: The design concept and the dosimetric characteristics of an applicator system for intraoperative radiation therapy (IORT) with special emphasis on alignment methods, the effect of a plastic scatterer in the beam, radiation leakage, and misalignment dosimetry, are presented in this paper. MATERIALS AND METHODS: A soft-docking system for a linear accelerator, which enables collimation of electron beams (4-22 MeV) for IORT has been developed. The system includes twenty-one circular polymethylmethacrylate (PMMA) treatment cones of different lengths, diameters and end angles. All in-water measurements are made using p-type silicon diode detectors. RESULTS: The effect of introducing a PMMA scatterer in the therapeutic beam includes increased surface dose values (above 83% for all nominal electron energies and for all cones) and improved dose homogeneity within the therapeutic range. Electrons scattered from the inside wall of the cone result in dose profile horns at depth of dose maximum always lower than 109%. The radiation leakage outside the cone is less than 13%. Large changes in the dose profiles occur if the intraoperative cone is misaligned more than 0.5. CONCLUSION: The alignment procedure of the soft-docking system is easy to handle and the applicator design provides adequate collimation of electron beams for IORT.
  •  
6.
  • Björk, Peter, et al. (author)
  • Dosimetry characteristics of degraded electron beams investigated by Monte Carlo calculations in a setup for intraoperative radiation therapy.
  • 2002
  • In: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 47:2, s. 239-256
  • Journal article (peer-reviewed)abstract
    • Degraded electron beams, as used for intraoperative radiation therapy (IORT) or similar complicated dosimetric situations, have different characteristics compared to conventional electron therapy beams. If international dosimetry protocols are applied in a direct manner to such degraded beams, uncertainties will be introduced in the absorbed dose determination. The Monte Carlo method has been used to verify experimentally determined relative absorbed dose distributions and output factors in an IORT geometry. Monte Carlo generated dose distributions are mostly within +/-2% or +/-2 mm of measured data. The simulated output variation between the IORT cones (relative output factors) are mostly within 2% of measured values. By comparing IORT and conventional electron beam characteristics (e.g. energy spectra, angular distributions and the contributions of different system components to these quantities) limitations and uncertainties of commonly used dosimetric techniques in IORT electron fields are quantified. The intraoperative treatment field contains a larger amount of scattered electrons, which leads to a broader energy spectrum as well as a wider angular distribution of electrons at the phantom surface. The dose from the scattered electrons can contribute up to 40% of the total dose at a depth of dose maximum, compared to approximately 10% for standard beams. A study of the energy spectra at the reference depth reveals that an uncertainty of the order of 1% can be introduced if ionization chamber based dosimetry is used to determine output factors for the investigated IORT system. We recommend that relative absorbed dose distributions and output factors in IORT electron beams and for similar complicated dosimetric situations should be determined with detectors having a small energy and angular dependence (e.g. diamond detectors or p-Si diodes).
  •  
7.
  •  
8.
  •  
9.
  •  
10.
  • Blad, Börje, et al. (author)
  • The influence of air humidity on an unsealed ionization chamber in a linear accelerator
  • 1996
  • In: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 41:11, s. 2541-2548
  • Journal article (peer-reviewed)abstract
    • The safe and accurate delivery of the prescribed absorbed dose is the central function of the dose monitoring and beam stabilization system in a medical linear accelerator. The absorbed dose delivered to the patient during radiotherapy is often monitored by a transmission ionization chamber. Therefore it is of utmost importance that the chamber behaves correctly. We have noticed that the sensitivity of an unsealed chamber in a Philips SL linear accelerator changes significantly, especially during and after the summer season. The reason for this is probably a corrosion effect of the conductive plates in the chamber due to the increased relative humidity during hot periods. We have found that the responses of the different ion chamber plates change with variations in air humidity and that they do not return to their original values when the air humidity is returned to ambient conditions.
  •  
11.
  • Bohsung, J, et al. (author)
  • IMRT treatment planning - A comparative inter-system and intor-centre planning exercise of the ESTRO QUASIMODO group
  • 2005
  • In: Radiotherapy and Oncology. - : Elsevier BV. - 1879-0887 .- 0167-8140. ; 76:3, s. 354-361
  • Journal article (peer-reviewed)abstract
    • Background and purpose: The purpose of this work was a comparison of realistic IMRT plans based on the same CT-image data set and a common predefined set of dose objectives for the planning target volume and the organs at risk. This work was part of the larger European QUASIMODO IMRT verification project. Materials and methods: Eleven IMRT. plans were produced by nine different European groups, each applying a representative set of. clinically used IMRT treatment planning systems. The plans produced were to be deliverable in a clinically acceptable treatment time with the local technical equipment. All plans were characterized using a set of different quality measures such as dose-volume histograms, number of monitor units and treatment time. Results: Only one plan was able to fulfil all dose objectives strictly; six plans failed some of the objectives but were still considered to be clinically acceptable; four plans were not able to reach the objectives. Additional quality scores such as the number of monitor units and treatment time showed large variations, which mainly depend on the delivery technique. Conclusion: The presented planning study showed that with nearly all presently available IMRT planning and delivery systems comparable dose distributions could be achieved if the planning goals are clearly defined in advance.
  •  
12.
  •  
13.
  • Brockstedt, Sara, et al. (author)
  • BE AWARE OF NEUTRONS OUTSIDE SHORT MAZES FROM 10-MV LINEAR ACCELERATORS X-RAYS IN RADIOTHERAPY FACILITIES.
  • 2015
  • In: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 165:1-4, s. 464-467
  • Journal article (peer-reviewed)abstract
    • During the radiation survey of a reinstalled 10-MV linear accelerator in an old radiation treatment facility, high dose rates of neutrons were observed. The area outside the maze entrance is used as a waiting room where patients, their relatives and staff other than those involved in the actual treatment can freely pass. High fluence rates of neutrons would cause an unnecessary high effective dose to the staff working in the vicinity of such a system, and it can be several orders higher than the doses received due to X-rays at the same location. However, the common knowledge appears to have been that the effect of neutrons at 10-MV X-ray linear accelerator facilities is negligible and shielding calculations models seldom mention neutrons for this operating energy level. Although data are scarce, reports regarding this phenomenon are now emerging. For the future, it is advocated that contributions from neutrons are considered already during the planning stage of new or modified facilities aimed for 10 MV and that estimated dose levels are verified.
  •  
14.
  • Ceberg, Crister, et al. (author)
  • Prediction of stopping-power ratios in flattening-filter free beams.
  • 2010
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 37:3, s. 1164-1168
  • Journal article (peer-reviewed)abstract
    • PURPOSE: In recent years, there has been an increasing interest in flattening-filter free (FFF) beams. However, since the removal of the flattening filter will affect both the mean and the variance of the energy spectrum, current beam-quality specifiers may not be adequate for reference dosimetry in such beams. The purpose of this work was to investigate an alternative, more general beam-quality specifier. METHODS: The beam-quality specifier used in this work was a combination of the kerma-weighted mean and the coefficient of variation of the linear attenuation coefficient in water. These parameters can in theory be determined from narrow-beam transmission measurements using a miniphantom "in-air," which is a measurement condition well suited also to small and nonstandard fields. The relation between the Spencer-Attix stopping-power ratios and this novel beam-quality specifier was described by a simple polynomial. For reference, the authors used Monte Carlo calculated spectra and stopping-power data for nine different beams, with and without flattening filter. RESULTS: The polynomial coefficients were obtained by least-squares optimization. For all beams included in this investigation, the average of the differences between the predicted and the Monte Carlo calculated stopping-power ratios was 0.02 +/- 0.17% (1 SD) (including TomoTherapy and CyberKnife example beams). CONCLUSIONS: An alternative dual-parameter beam-quality specifier was investigated. The evaluation suggests that it can be used successfully to predict stopping-power ratios in FFF as well as conventional beams, regardless of filtration.
  •  
15.
  • Cunningham, Joanne, et al. (author)
  • Radiation Oncology Safety Information System (ROSIS) - Profiles of participants and the first 1074 incident reports
  • 2010
  • In: Radiotherapy and Oncology. - : Elsevier BV. - 1879-0887 .- 0167-8140. ; 97:3, s. 601-607
  • Journal article (peer-reviewed)abstract
    • Background and purpose: The Radiation Oncology Safety Information System (ROSIS) was established in 2001. The aim of ROSIS is to collate and share information on incidents and near-incidents in radiotherapy, and to learn from these incidents in the context of departmental infrastructure and procedures. Materials and methods: A voluntary web-based cross-organisational and international reporting and learning system was developed (cf. the www.rosis.info website). Data is collected via online Department Description and Incident Report Forms. A total of 101 departments, and 1074 incident reports are reviewed. Results: The ROSIS departments represent about 150,000 patients, 343 megavoltage (MV) units, and 114 brachytherapy units. On average, there are 437 patients per MV unit, 281 per radiation oncologist, 387 per physicist and 353 per radiation therapy technologist (RT/RTT). Only 14 departments have a completely networked system of electronic data transfer, while 10 departments have no electronic data transfer. On average seven quality assurance (QA) or quality control (QC) methods are used at each department. A total of 1074 ROSIS reports are analysed; 97.7% relate to external beam radiation treatment and 50% resulted in incorrect irradiation. Many incidents arise during pre-treatment but are not detected until later in the treatment process. Where an incident is not detected prior to treatment, an average of 22% of the prescribed treatment fractions were delivered incorrectly. The most commonly reported detection methods were "found at time of patient treatment" and during "chart-check". Conclusion: While the majority of the incidents that reported to this international cross-organisational reporting system are of minor dosimetric consequence, they affect on average more than 20% of the patient's treatment fractions. Nonetheless, defence-in-depth is apparent in departments registered with ROSIS. This indicates a need for further evaluation of the effectiveness of quality controls. (C) 2010 Published by Elsevier Ireland Ltd. Radiotherapy and Oncology 97 (2010) 601-607
  •  
16.
  • Dalaryd, Mårten, et al. (author)
  • A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements.
  • 2010
  • In: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 55:23, s. 7333-7344
  • Journal article (peer-reviewed)abstract
    • A Monte Carlo model of an Elekta Precise linear accelerator has been built and verified by measured data for a 6 and 10 MV photon beam running with and without a flattening filter in the beam line. In this study the flattening filter was replaced with a 6 mm thick copper plate, provided by the linac vendor, in order to stabilize the beam. Several studies have shown that removal of the filter improves some properties of the photon beam, which could be beneficial for radiotherapy treatments. The investigated characteristics of this new beam included output, spectra, mean energy, half value layer and the origin of scattered photons. The results showed an increased dose output per initial electron at the central axis of 1.76 and 2.66 for the 6 and 10 MV beams, respectively. The number of scattered photons from the accelerator head was reduced by (31.7 ± 0.03)% (1 SD) for the 6 MV beam and (47.6 ± 0.02)% for the 10 MV beam. The photon energy spectrum of the unflattened beam was softer compared to a conventional beam and did not vary significantly with the off-axis distance, even for the largest field size (0-20 cm off-axis).
  •  
17.
  • Dalaryd, Mårten, et al. (author)
  • Combining tissue-phantom ratios to provide a beam-quality specifier for flattening filter free photon beams.
  • 2014
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 41:11
  • Journal article (peer-reviewed)abstract
    • There are currently several commercially available radiotherapy treatment units without a flattening filter in the beam line. Unflattened photon beams have an energy and lateral fluence distribution that is different from conventional beams and, thus, their attenuation properties differ. As a consequence, for flattening filter free (FFF) beams, the relationship between the beam-quality specifier TPR20,10 and the Spencer-Attix restricted water-to-air mass collision stopping-power ratios, L̄/ρair (water), may have to be refined in order to be used with equivalent accuracy as for beams with a flattening filter. The purpose of this work was twofold. First, to study the relationship between TPR20,10 and L̄/ρair (water) for FFF beams, where the flattening filter has been replaced by a metal plate as in most clinical FFF beams. Second, to investigate the potential of increasing the accuracy in determining L̄/ρair (water) by adding another beam-quality metric, TPR10,5. The relationship between L̄/ρair (water) and %dd(10)x for beams with and without a flattening filter was also included in this study.
  •  
18.
  • Fogliata, A., et al. (author)
  • Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy
  • 2012
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 39:10, s. 6455-6464
  • Journal article (peer-reviewed)abstract
    • Purpose: Flattening filter free (FFF) beams generated by medical linear accelerators have recently started to be used in radiotherapy clinical practice. Such beams present fundamental differences with respect to the standard filter flattened (FF) beams, making the generally used dosimetric parameters and definitions not always viable. The present study will propose possible definitions and suggestions for some dosimetric parameters for use in quality assurance of FFF beams generated by medical linacs in radiotherapy. Methods: The main characteristics of the photon beams have been analyzed using specific data generated by a Varian TrueBeam linac having both FFF and FF beams of 6 and 10 MV energy, respectively. Results: Definitions for dose profile parameters are suggested starting from the renormalization of the with respect to the corresponding FF beam. From this point the flatness concept has been translated into one of "unflatness" and other definitions have been proposed, maintaining a strict parallelism between FFF and FF parameter concepts. Conclusions: Ideas for quality controls used in establishing a quality assurance program when introducing FFF, beams into the clinical environment are given here, keeping them similar to those used for standard FF beams. By following the suggestions in this report, the authors foresee that the introduction of FFF beams into a clinical radiotherapy environment will be as safe and well controlled as standard beam modalities using the existing guidelines. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4754799]
  •  
19.
  • Fogliata, Antonella, et al. (author)
  • On the dosimetric behaviour of photon dose calculation algorithms in the presence of simple geometric heterogeneities: comparison with Monte Carlo calculations
  • 2007
  • In: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 52:5, s. 1363-1385
  • Journal article (peer-reviewed)abstract
    • A comparative study was performed to reveal differences and relative figures of merit of seven different calculation algorithms for photon beams when applied to inhomogeneous media. The following algorithms were investigated: Varian Eclipse: the anisotropic analytical algorithm, and the pencil beam with modified Batho correction; Nucletron Helax-TMS: the collapsed cone and the pencil beam with equivalent path length correction; CMS XiO: the multigrid superposition and the fast Fourier transform convolution; Philips Pinnacle: the collapsed cone. Monte Carlo simulations ( MC) performed with the EGSnrc codes BEAMnrc and DOSxyznrc from NRCC in Ottawa were used as a benchmark. The study was carried out in simple geometrical water phantoms(rho = 1.00 g cm(-3)) with inserts of different densities simulating light lung tissue (rho = 0.035 g cm(-3)), normal lung (rho = 0.20 g cm(-3)) and cortical bone tissue (rho = 1.80 g cm(-3)). Experiments were performed for low- and high-energy photon beams (6 and 15 MV) and for square (13 x 13 cm(2)) and elongated rectangular (2.8 x 13 cm(2)) fields. Analysis was carried out on the basis of depth dose curves and transverse profiles at several depths. Assuming the MC data as reference,. index analysis was carried out distinguishing between regions inside the non-water inserts or inside the uniformwater. For this study, a distance to agreement was set to 3 mm while the dose difference varied from 2% to 10%. In general all algorithms based on pencil-beam convolutions showed a systematic deficiency in managing the presence of heterogeneous media. In contrast, complicated patterns were observed for the advanced algorithmswith significant discrepancies observed between algorithms in the lighter materials (rho = 0.035 g cm(-3)), enhanced for the most energetic beam. For denser, and more clinical, densities a better agreement among the sophisticated algorithms with respect to MC was observed.
  •  
20.
  • Georg, Dietmar, et al. (author)
  • Current status and future perspective of flattening filter free photon beams
  • 2011
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 38:3, s. 1280-1293
  • Journal article (peer-reviewed)abstract
    • Purpose: Flattening filters (FFs) have been considered as an integral part of the treatment head of a medical accelerator for more than 50 years. The reasons for the longstanding use are, however, historical ones. Advanced treatment techniques, such as stereotactic radiotherapy or intensity modulated radiotherapy have stimulated the interest in operating linear accelerators in a flattening filter free (FFF) mode. The current manuscript reviews treatment head physics of FFF beams, describes their characteristics and the resulting potential advantages in their medical use, and closes with an outlook. Methods: A number of dosimetric benefits have been determined for FFF beams, which range from increased dose rate and dose per pulse to favorable output ratio in-air variation with field size, reduced energy variation across the beam, and reduced leakage and out-of-field dose, respectively. Finally, the softer photon spectrum of unflattened beams has implications on imaging strategies and radiation protection. Results: The dosimetric characteristics of FFF beams have an effect on treatment delivery, patient comfort, dose calculation accuracy, beam matching, absorbed dose determination, treatment planning, machine specific quality assurance, imaging, and radiation protection. When considering conventional C-arm linacs in a FFF mode, more studies are needed to specify and quantify the clinical advantages, especially with respect to treatment plan quality and quality assurance. Conclusions: New treatment units are already on the market that operate without a FF or can be operated in a dedicated clinical FFF mode. Due to the convincing arguments of removing the FF, it is expected that more vendors will offer dedicated treatment units for advanced photon beam therapy in the near future. Several aspects related to standardization, dosimetry, treatment planning, and optimization need to be addressed in more detail in order to facilitate the clinical implementation of unflattened beams. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3554643]
  •  
21.
  • Georg, Dietmar, et al. (author)
  • Photon beam quality variations of a flattening filter free linear accelerator
  • 2010
  • In: Medical Physics. - : Wiley. - 0094-2405. ; 37:1, s. 49-53
  • Journal article (peer-reviewed)abstract
    • Purpose: Recently, there has been an increasing interest in operating conventional linear accelerators without a flattening filter. The aim of this study was to determine beam quality variations as a function of off-axis ray angle for unflattened beams. In addition, a comparison was made with the off-axis energy variation in flattened beams. Methods: Two Elekta Precise linear accelerators were modified in order to enable radiation delivery with and without the flattening filter in the beam line. At the Medical University Vienna (Vienna, Austria), half value layer (HVL) measurements were performed for 6 and 10 MV with an in-house developed device that can be easily mounted on the gantry. At St. Luke's Hospital (Dublin, Ireland), measurements were performed at 6 MV in narrow beam geometry with the gantry tilted around 270 degrees with pinhole collimators, an attenuator, and the chamber positioned on the table. All attenuation measurements were performed with ionization chambers and a buildup cap (2 mm brass) or a PMMA mini phantom (diameter 3 cm, measurement depth 2.5 cm). Results: For flattened 6 and 10 MV photon beams from the Elekta linac the relative HVL(theta) varies by about 11% for an off-axis ray angle theta=10 degrees. These results agree within +/-2% with a previously proposed generic off-axis energy correction. For unflattened beams, the variation was less than 5% in the whole range of off-axis ray angles up to 10 degrees. The difference in relative HVL data was less than 1% for unflattened beams at 6 and 10 MV. Conclusions: Off-axis energy variation is rather small in unflattened beams and less than half the one for flattened beams. Thus, ignoring the effect of off-axis energy variation for dose calculations in unflattened beams can be clinically justified. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3264617]
  •  
22.
  • Hansson, H, et al. (author)
  • Verification of a pencil beam based treatment planning system: output factors for open photon beams shaped with MLC or blocks
  • 1999
  • In: Physics in Medicine and Biology. - 1361-6560. ; 44:9, s. 201-207
  • Journal article (peer-reviewed)abstract
    • The accuracy of monitor unit calculations from a pencil beam based, three-dimensional treatment planning system (3D TPS) has been evaluated for open irregularly shaped photon fields. The dose per monitor unit was measured in water and in air for x-ray beam qualities from 6 to 15 MV. The fields were shaped either with a multileaf collimator (MLC) or with customized alloy blocks. Calculations from the 3D TPS were compared with measurements. The agreement between calculated and measured dose per monitor unit depended on field size and the amount of blocking and was within 3% for the MLC-shaped fields. The deviation could be traced to limitations in head scatter modelling for the MLC. For fields shaped with alloy blocks, the dose per monitor unit was calculated to be within 1.6% of measured values for all fields studied. The measured and calculated relative phantom scatter for fields with the same equivalent field size were identical for MLC and alloy shaped fields. These results indicate that the accuracy in the TPS calculations for open irregular fields, shaped with MLC or blocks, is satisfactory for clinical situations.
  •  
23.
  • Hurkmans, Coen, et al. (author)
  • Dosimetric verification of open asymmetric photon fields calculated with a treatment planning system based on dose-to-energy-fluence concepts
  • 1996
  • In: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 41:8, s. 1277-1290
  • Journal article (peer-reviewed)abstract
    • Output normalized dose profiles for asymmetric open photon fields has been calculated using a commercial treatment planning system (TPS) based on a dose-to-energy-fluence concept. The model does not require any additional measurements for off-axis fields. Calculations are compared with measurements for quadratic fields of 5 cm x 5 cm up to 20 cm x 20 cm, with their geometric field centre positioned 10 cm off-axis in the in-plane direction. The measurements include depth doses and profiles in-plane as well as cross-plane for nominal photon energies of 4, 6 and 18 MV x-rays. Both calculated and measured doses are normalized with respect to a 10 cm x 10 cm reference field, therefore making it possible to compare not only the relative distributions but also the absolute dose levels; that is, calculation of monitor units is included. The calculated depth-dose curves are generally in good agreement with measured data with an accuracy at the absolute dose level of 2% at depths beyond the dose maximum. The cross-plane profiles are calculated with an accuracy better than 3% within the field. The 'tilt' towards the collimator central axis of the in-plane profiles is predicted by the model, but is somewhat overestimated at large depths. The system provides the possibility to separate the primary and scattered parts of the dose and the cause of this tilting was studied by comparing calculated phantom-scattering and head-scattering dose profiles for a symmetric 40 cm x 20 cm field to dose profiles for an asymmetric 20 cm x 20 cm field. The tilting is shown to originate from a change both in phantom scattering and in head scattering compared to the case of symmetrical fields. The results indicate that the investigated TPS can calculate dose distributions in open asymmetric fields with a high degree of accuracy, typically better than 2-3%.
  •  
24.
  • Hurkmans, Coen, et al. (author)
  • Limitations of a pencil beam approach to photon dose calculations in the head and neck region
  • 1996
  • In: Medical Dosimetry. - : Elsevier BV. - 1873-4022 .- 0958-3947. ; 21:1, s. 38-38
  • Journal article (peer-reviewed)abstract
    • The inherent limitations of a specific pencil beam model have been studied when applied to a cylindrical geometry simulating the neck region. A comparison is made between measured and calculated absorbed dose in a cylindrical phantom. The goal is to quantify the deviations in the absorbed dose level, i.e., the dose per monitor unit, when photons are used for the treatment of head and neck tumours. Square fields ranging from 5 x 5 up to 30 x 30 cm[super:2] are studied for photon beam energies of [super:60]Co, 4, 6 and 18 MV. Ionisation chamber measurements have been performed in the cylinder as well as in two other configurations in order to trace the origin of possible deviations. For 18 MV no significant deviations are found between measurement and calculation in the cylindrical configuration. For the lower energies, an over-estimation of the calculated dose in the cylindrical configuration up to about 6% for a 20 x 20-cm[super:2] [super:60]Co field has been found. These deviations have been traced to the basic approximation for the integration volume for phantom scatter calculations inherent in this pencil beam implementation.
  •  
25.
  • Hurkmans, Coen, et al. (author)
  • Limitations of a pencil beam approach to photon dose calculations in the head and neck region
  • 1995
  • In: Radiotherapy and Oncology. - 1879-0887. ; 37:1, s. 74-80
  • Journal article (peer-reviewed)abstract
    • The inherent limitations of a specific pencil beam model have been studied when applied to a cylindrical geometry simulating the neck region. A comparison is made between measured and calculated absorbed dose in a cylindrical phantom. The goal is to quantify the deviations in the absorbed dose level, i.e., the dose per monitor unit, when photons are used for the treatment of head and neck tumours. Square fields ranging from 5 x 5 up to 30 x 30 cm2 are studied for photon beam energies of 60Co, 4, 6 and 18 MV. Ionisation chamber measurements have been performed in the cylinder as well as in two other configurations in order to trace the origin of possible deviations. For 18 MV no significant deviations are found between measurement and calculation in the cylindrical configuration. For the lower energies, an overestimation of the calculated dose in the cylindrical configuration up to about 6% for a 20 x 20-cm2 60Co field has been found. These deviations have been traced to the basic approximation for the integration volume for phantom scatter calculations inherent in this pencil beam implementation.
  •  
Skapa referenser, mejla, bekava och länka
  • Result 1-25 of 67

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view