Observations on the Failure of Stent-Grafts in the Aortic Arch.
- 2007
-
In: European Journal of Vascular and Endovascular Surgery. - : Elsevier BV. - 1532-2165 .- 1078-5884. ; 34:4, s. 6-451
-
Journal article (peer-reviewed)abstract
- Abstract INTRODUCTION: The results of endovascular stent-grafts in the abdominal aorta and descending thoracic aorta have been encouraging. Expanding the use of thoracic stent-grafts in to the aortic arch has been associated with increasing numbers of complications. Recently isolated cases of stent-graft collapse have been reported. METHODS: This was a multi-centre European case series. Data was collected retrospectively on seven patients from five experienced endovascular centres with thoracic stent-graft collapse. RESULTS: Of the seven patients four were treated for traumatic aortic rupture. Six were male, median age 33 (range 17-54) years. During the ensuing 2 months all patients suffered stent-graft collapse. This was symptomatic in 3 patients and the rest were identified on CT. Endovascular management was possible in 6/7 patients using either a balloon expandable stent (n=6) or further stent-graft (n=1). Two patients had persistent type I endoleak despite treatment. Two of the 7 patients died, both of which presented with symptomatic thoracic stent-graft occlusion. Both deaths were a direct result of stent-graft collapse. CONCLUSIONS: Thoracic stent-graft collapse may be asymptomatic underscoring the importance of stent-graft surveillance. Endovascular management of collapse is possible in most cases using a large balloon expandable stent. Symptomatic collapse is associated with high morbidity and mortality.
|
|