Microdosimetry for a carbon ion beam using track-etched detectors
- 2015
-
In: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 0144-8420 .- 1742-3406. ; 166:1-4, s. 247-252
-
Journal article (peer-reviewed)abstract
- Track-etched detectors (TED) have been used as linear energy transfer (LET) spectrometers in heavy ion beams for many years. LET spectra and depth-dose distribution of a carbon ion beam were measured behind polymethylmethacrylate degraders at Heavy Ion Medical Accelerator in Chiba, Japan. The measurements were performed along monoenergetic beam with energy 290 MeV u-1 in different positions: (1) at beam extraction area, (2) at beginning, (3) maximum and (4) behind the Bragg peak region (0, 117, 147 and 151 mm of water-equivalent depth, respectively). The LET spectra inside and outside of the primary ion beam have been evaluated. TED record only heavy charged particles with LET above 8-10 keV μm-1, while electrons and ions with lower LET are not detected. The Geant4 simulation toolkit version 4.9.6.P01 has been used to estimate the contribution of non-detected particles to absorbed dose. Presented results demonstrate the applicability of TED for microdosimetry measurements in therapeutic carbon ion beams.
|
|