Assessing wear of the acetabular cup using computed tomography : an ex vivo study
-
- Olivecrona, Lotta (author)
- Karolinska Institutet, Department of Clinical Science, Intervention and Technology, Division of Radiology
-
- Jedenmalm, Anneli (author)
- KTH,Materialvetenskap
-
- Aspelin, Peter (author)
- Karolinska Institutet
-
show more...
-
- Weidenhielm, Lars (author)
- Karolinska Institutet
-
- Noz, Marilyn E. (author)
- New York University, Department of Radiology,Nuclear Medicine
-
- Maguire Jr., Gerald Q. (author)
- KTH,Kommunikationssystem, CoS,Computer Communications
-
- Zeleznik, Michael P. (author)
- RAHD Oncology Products, St. Louis. Mo
-
- Olivecrona, Henrik (author)
- Department of Hand Surgery, Södersjukhuset, Stockholm
-
show less...
-
(creator_code:org_t)
- 2016-08-30
- 2005
- English.
-
In: Acta Radiologica. - London, UK : Informa Healthcare. - 0284-1851 .- 1600-0455. ; 46:8, s. 852-857
- Related links:
-
https://urn.kb.se/re...
-
show more...
-
https://doi.org/10.1...
-
http://kipublication...
-
show less...
Abstract
Subject headings
Close
- Purpose: To validate a clinically useful method for measuring acetabular cup wear using computed tomography (CT). Material and Methods: Eight uncemented acetabular cups were scanned twice ex vivo using CT. The linear penetration depth of the femoral component head into the cup and the thickness of the remaining polyethylene liner were measured in the CT volumes using dedicated software. Two independent examiners twice assessed each volume. The CT measurements were compared to direct measurements using a coordinate measuring device and micrometer measurements. Results: Accuracy of wear measurements expressed as penetration depth was +/- 0.6 and +/- 1.0 mm for the two examiners, respectively, with no significant differences between examiners, trials, and CT scans. Accuracy of measurements of remaining polyethylene was +/- 1.3 and +/- 1.0 mm, respectively, for the two examiners. Systematic differences between examiners were found, but no significant differences between trials and CT scans. These differences were due to different interpretations of metal artifacts in the volumes. Conclusion: The proposed CT method for evaluating wear as head penetration depth allows for reliable wear detection at a clinically relevant level. Measurements of remaining polyethylene on CT volumes are not as reliable as wear measurements owing to metal artifacts.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Ortopedi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Orthopaedics (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Keyword
- Adults
- Computer applications-3D
- Hip
- CT
- Hip
- skeletal
- Hip
- spiral
- Technical aspects
- Medical engineering
- Medicinsk teknik
- SRA - ICT
- SRA - Informations- och kommunikationsteknik
Publication and Content Type
- ref (subject category)
- art (subject category)
Find in a library
To the university's database