SwePub
Sök i LIBRIS databas

  Extended search

onr:"swepub:oai:DiVA.org:kth-6941"
 

Search: onr:"swepub:oai:DiVA.org:kth-6941" > Physical characteri...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors

Åslund, Magnus (author)
KTH,Fysik
Cederström, Björn (author)
KTH,Fysik
Lundqvist, Mats (author)
KTH,Fysik
show more...
Danielsson, Mats (author)
KTH,Fysik
show less...
 (creator_code:org_t)
2007-05-08
2007
English.
In: Medical physics (Lancaster). - : Wiley. - 0094-2405. ; 34:6, s. 1918-1925
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • The physical performance of a scanning multislit full field digital mammography system was determined using basic image quality parameters. The system employs a direct detection detector comprised of linear silicon strip sensors in an edge-on geometry connected to photon counting electronics. The pixel size is 50 mu m and the field of view 24 x 26 cm(2). The performance was quantified using the presampled modulation transfer function, the normalized noise power spectrum and the detective quantum efficiency (DQE). Compared to conventional DQE methods, the scanning geometry with its intrinsic scatter rejection poses additional requirements on the measurement setup, which are investigated in this work. The DQE of the photon counting system was found to be independent of the dose level to the detector in the 7.6-206 mu Gy range. The peak DQE was 72% and 73% in the scan and slit direction, respectively, measured with a 28 kV W-0.5 mm Al anodefilter combination with an added 2 mm Al filtration.

Subject headings

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)

Keyword

mammography; detective quantum efficiency; scanning; MODULATION TRANSFER-FUNCTION; IMAGE QUALITY CHARACTERIZATION; QUANTUM EFFICIENCY; PERFORMANCE; ENERGY
Medical engineering
Medicinsk teknik

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Åslund, Magnus
Cederström, Björ ...
Lundqvist, Mats
Danielsson, Mats
About the subject
ENGINEERING AND TECHNOLOGY
ENGINEERING AND ...
and Medical Engineer ...
and Medical Laborato ...
Articles in the publication
Medical physics ...
By the university
Royal Institute of Technology

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view