Overestimation of flow velocity through leaks in mechanical valve prostheses and through small orifices by continuous-wave Doppler.

Flachskampf, Frank A (author)
Institute of Aerodynamics RWTH, Aachen, Germany
Köhler, Josef (author)
Institute of Aerodynamics RWTH, Aachen, Germany
Ask, Per (author)
Linköpings universitet,Fysiologisk mätteknik,Tekniska högskolan
show more...
Lepper, Wolfgang (author)
Institute of Aerodynamics RWTH, Aachen, Germany
Hanrath, Peter (author)
Institute of Aerodynamics RWTH, Aachen, Germany
show less...
 (creator_code:org_t)
1997
1997
English.
In: Journal of the American Society of Echocardiography. - 0894-7317 .- 1097-6795. ; 10:9, s. 904-914
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • The reliability of continuous-wave Doppler flow velocity measurements through small regurgitant lesions, such as in prosthetic leakage, has not been systematically analyzed. To evaluate the accuracy of continuous-wave Doppler in prosthetic valve leakage and small orifices in an in vitro, steady-flow model-flow velocities through the leaks of twelve intact mechanical prostheses and through six circular nozzles (area 0.5 to 20 mm2) were measured at pressure drops between 30 and 105 mm Hg. These results were compared with those predicted by the modified Bernoulli equation. Laser Doppler anemometry of flow velocities through the nozzles was also performed. Despite high correlation, there was substantial overestimation of Bernoulli predicted velocities by echo Doppler in the prosthetic leaks (mean +12.3% +/- 9.4%; range 90.3% to 143.4%). In the nozzles < or = 10 mm2, but not in the largest (20 mm2) nozzle, there was also overestimation of the Bernoulli predicted velocities (mean +6.2% +/- 2%). Laser Doppler anemometry of flow velocities through the nozzles showed slightly lower values than predicted by the Bernoulli equation. Thus, continuous-wave echo Doppler overestimates flow velocities through small orifices. This apparently is, at least in part, due to transit time effects and should be taken into account when using echo Doppler in small (< 10 mm2) orifices, such as in mild to moderate regurgitant lesions and prosthetic valve leakage.

Subject headings

TEKNIK OCH TEKNOLOGIER  -- Maskinteknik -- Strömningsmekanik och akustik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Mechanical Engineering -- Fluid Mechanics and Acoustics (hsv//eng)

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Flachskampf, Fra ...
Köhler, Josef
Ask, Per
Lepper, Wolfgang
Hanrath, Peter
About the subject
ENGINEERING AND TECHNOLOGY
ENGINEERING AND ...
and Mechanical Engin ...
and Fluid Mechanics ...
Articles in the publication
Journal of the A ...
By the university
Linköping University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view