Transmission ionization chambers for measurements of air collision kerma integrated over beam area. Factors limiting the accuracy of calibration

Larsson, Peter (author)
Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet
Persliden, Jan (author)
Östergötlands Läns Landsting,Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet,Radiofysikavdelningen
Sandborg, Michael (author)
Östergötlands Läns Landsting,Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet,Radiofysikavdelningen
show more...
Alm Carlsson, Gudrun (author)
Östergötlands Läns Landsting,Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet,Radiofysikavdelningen
show less...
 (creator_code:org_t)
1999-01-01
1996
English.
In: Physics in Medicine and Biology. - : IOP Publishing. - 0031-9155 .- 1361-6560. ; 41:11, s. 2381-2398
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Kerma - area product meters (KAP meters) are frequently used in diagnostic radiology to measure the integral of air-collision kerma over an area perpendicular to the x-ray beam. In this work, a precise method for calibrating a KAP meter to measure is described and calibration factors determined for a broad range of tube potentials (40 - 200 kV). The integral is determined using a large number of TL dosimeters spread over and outside the nominal field area defined as the area within 50% of maximum . The method is compared to a simplified calibration method which approximates the integral by multiplying the kerma in the centre of the field by the nominal field area . While the calibration factor using the precise method is independent of field area and distance from the source, that using the simplified method depends on both. This can be accounted for by field inhomogeneities caused by the heel effect, extrafocal radiation and scattered radiation from the KAP meter. The deviations between the calibration factors were as large as for collimator apertures of and distances from the source of 50 - 160 cm. The uncertainty in the calibration factor using the precise method was carefully evaluated and the expanded relative uncertainty estimated to be with a confidence level of 95%.

Keyword

MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view