ACCURACY OF CT NUMBERS OBTAINED BY DIRA AND MONOENERGETIC PLUS ALGORITHMS IN DUAL-ENERGY COMPUTED TOMOGRAPHY
-
- Magnusson, Maria, 1961- (author)
- Linköpings universitet,Datorseende,Tekniska fakulteten,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för diagnostik och specialistmedicin
-
- Sandborg, Michael, 1961- (author)
- Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Medicinsk strålningsfysik
-
- Alm Carlsson, Gudrun, 1939- (author)
- Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
-
show more...
-
- Henriksson, Lilian (author)
- Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Region Östergötland, Röntgenkliniken i Linköping
-
- Carlsson Tedgren, Åsa, 1968- (author)
- Karolinska Institutet,Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Department of Medical Radiation Physics and Nuclear Medicine; Karolinska University Hospital , Stockholm, Sweden
-
- Malusek, Alexandr, 1968- (author)
- Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
-
show less...
-
(creator_code:org_t)
- 2021-07-16
- 2021
- English.
-
In: Radiation Protection Dosimetry. - : Oxford University Press. - 0144-8420 .- 1742-3406. ; 195:3-4, s. 212-217
- Related links:
-
https://liu.diva-por... (primary) (Raw object)
-
show more...
-
https://academic.oup...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
http://kipublication...
-
show less...
Abstract
Subject headings
Close
- Dual-energy computed tomography (CT) can be used in radiotherapy treatment planning for the calculation of absorbed dose distributions. The aim of this work is to evaluate whether there is room for improvement in the accuracy of the Monoenergetic Plus algorithm by Siemens Healthineers. A Siemens SOMATOM Force scanner was used to scan a cylindrical polymethyl methacrylate phantom with four rod-inserts made of different materials. Images were reconstructed using ADMIRE and processed with Monoenergetic Plus. The resulting CT numbers were compared with tabulated values and values simulated by the proof-of-a-concept algorithm DIRA developed by the authors. Both the Monoenergetic Plus and DIRA algorithms performed well; the accuracy of attenuation coefficients was better than about ±1% at the energy of 70 keV. Compared with DIRA, the worse performance of Monoenergetic Plus was caused by its (i) two-material decomposition to iodine and water and (ii) imperfect suppression of the beam hardening artifact in ADMIRE.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)