Search: onr:"swepub:oai:DiVA.org:liu-26776" >
Radiation dose opti...
Radiation dose optimization in coronary angiography and percutaneous coronary intervention (PCI) : II. Clinical evaluation
-
- Geijer, Håkan (author)
- Department of Radiology, Örebro Medical Centre Hospital, Örebro, Sweden
-
- Beckman, Karl-Wilhelm (author)
- Department of Medical Physics, Örebro Medical Centre Hospital, Örebro, Sweden
-
- Andersson, Tony, 1973- (author)
- Department of Radiology, Örebro Medical Centre Hospital, Örebro, Sweden
-
show more...
-
- Persliden, Jan, 1950- (author)
- Department of Medical Physics, Örebro Medical Centre Hospital, Örebro, Sweden
-
show less...
-
(creator_code:org_t)
- Springer Science and Business Media LLC, 2002
- 2002
- English.
-
In: European Radiology. - : Springer Science and Business Media LLC. - 0938-7994 .- 1432-1084. ; 12:11, s. 2813-2819
- Related links:
-
https://urn.kb.se/re...
-
show more...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- In a previous part of this study, the fluoroscopy dose rate was reduced in a cardiac catheterization laboratory. The objectives of the present study were to evaluate the effects in a clinical population undergoing percutaneous coronary intervention (PCI) of the dose-reducing measures detailed previously. Kerma area-product (KAP) values were first recorded for 154 patients undergoing PCI. Then, the fluoroscopy KAP rate was reduced from 44 to 16 mGy cm2/s by increasing filtration and reducing the image intensifier dose request. After this optimization, KAP was recorded for another 138 PCI procedures. After adjustment for differing proportions of combined procedures (coronary angiography+PCI), the total KAP was reduced to 67% of the original value with a 95% confidence interval from 57 to 78%, statistically significant. The mean total KAP values were 93.6 Gy cm2 before and 69.1 Gy cm2 after optimization. The KAP for digital acquisition did not change significantly. It is possible to make a large dose reduction in PCI by reducing the fluoroscopy dose rate. This dose reduction is beneficial for both patients and staff.
Keyword
- MEDICINE
- MEDICIN
Publication and Content Type
- ref (subject category)
- art (subject category)
Find in a library
To the university's database