In Vitro Assessment of Flow Patterns and Turbulence Intensity in Prosthetic Heart Valves Using Generalized Phase-Contrast MRI

Escobar Kvitting, John-Peder (author)
Östergötlands Läns Landsting,Linköpings universitet,Fysiologi,Hälsouniversitetet,Thorax-kärlkliniken
Dyverfeldt, Petter (author)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken
Sigfridsson, Andreas (author)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken
show more...
Franzen, Stefan (author)
Östergötlands Läns Landsting,Linköpings universitet,Thoraxkirurgi,Hälsouniversitetet,Thorax-kärlkliniken
Wigström, Lars (author)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken
Bolger, Ann F (author)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken
Ebbers, Tino (author)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken
show less...
 (creator_code:org_t)
John Wiley and Sons, Ltd, 2010
2010
English.
In: JOURNAL OF MAGNETIC RESONANCE IMAGING. - : John Wiley and Sons, Ltd. - 1053-1807 .- 1522-2586. ; 31:5, s. 1075-1080
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Purpose: To assess in vitro the three-dimensional mean velocity field and the extent and degree of turbulence intensity (TI) in different prosthetic heart valves using a generalization of phase-contrast MRI (PC-MRI). Materials and Methods: Four 27-mm aortic valves (Bjork-Shiley Monostrut tilting-disc, St. Jude Medical Standard bileaflet, Medtronic Mosaic stented and Freestyle stentless porcine valve) were tested under steady inflow conditions in a Plexiglas phantom. Three-dimensional PC-MRI data were acquired to measure the mean velocity field and the turbulent kinetic energy (TKE), a direction-independent measure of TI. Results: Velocity and TI estimates could be obtained up and downstream of the valves, except where metallic structure in the valves caused signal void. Distinct differences in the location, extent, and peak values of velocity and TI were observed between the valves tested. The maximum values of TKE varied between the different valves: tilting disc, 100 J/m(3); bileaflet, 115 J/m(3); stented, 200 J/m(3); stentless, 145 J/m(3). Conclusion: The TI downstream from a prosthetic heart valve is dependent on the specific valve design. Generalized PC-MRI can be used to quantify velocity and TI downstream from prosthetic heart valves, which may allow assessment of these aspects of prosthetic valvular function in postoperative patients.

Keyword

turbulence intensity
prosthetic heart valves
phase-contrast magnetic resonance imaging
MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view