SwePub
Sök i LIBRIS databas

  Extended search

onr:"swepub:oai:DiVA.org:oru-56419"
 

Search: onr:"swepub:oai:DiVA.org:oru-56419" > A visual grading st...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

A visual grading study for different administered activity levels in bone scintigraphy

Gustafsson, Agneta (author)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Radiofysikavdelningen US
Karlsson, Henrik (author)
Department of Radiation Physics, Kalmar County Hospital, Kalmar, Sweden
Olsson, Anna (author)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
show more...
Geijer, Håkan, 1961- (author)
Region Örebro län,Department of Radiology,Örebro University Hospital, Sweden
Olsson, Anna (author)
Department of Medical Physics, IMH, Linköping University, Linköping, Sweden; Radiation Physics UHL, County council of Östergötland, Linköping, Sweden; Center for Medical Image Science and Visualization, Linköping University, Linköping, Sweden
show less...
 (creator_code:org_t)
2014-05-04
2015
English.
In: Clinical Physiology and Functional Imaging. - : Wiley-Blackwell. - 1475-0961 .- 1475-097X. ; 35:3, s. 231-236
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Introduction: The aim of the study is to assess the administered activity levels versus visual-based image quality using visual grading regression (VGR) including an assessment of the newly stated image criteria for whole-body bone scintigraphy.Materials and methods: A total of 90 patients was included and grouped in three levels of administered activity: 400, 500 and 600 MBq. Six clinical image criteria regarding image quality was formulated by experienced nuclear medicine physicians. Visual grading was performed in all images, where three physicians rated the fulfilment of the image criteria on a four-step ordinal scale. The results were analysed using VGR. A count analysis was also made where the total number of counts in both views was registered.Results: The administered activity of 600 MBq gives significantly better image quality than 400 MBq in five of six criteria (P<005). Comparing the administered activity of 600 MBq to 500 MBq, four criteria of six show significantly better image quality (P<005). The administered activity of 500 MBq gives no significantly better image quality than 400 Mbq (P<005). The count analysis shows that none of the three levels of administrated activity fulfil the recommendations by the EANM.Conclusion: There was a significant improvement in perceived image quality using an activity level of 600 MBq compared to lower activity levels in whole-body bone scintigraphy for the gamma camera equipment end set-up used in this study. This type of visual-based grading study seems to be a valuable tool and easy to implement in the clinical environment.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinska och farmaceutiska grundvetenskaper -- Fysiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Basic Medicine -- Physiology (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)

Keyword

bone and bones
evaluation
radionuclide imaging
technetium Tc 99 m

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Gustafsson, Agne ...
Karlsson, Henrik
Olsson, Anna
Geijer, Håkan, 1 ...
About the subject
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Basic Medicine
and Physiology
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Clinical Medicin ...
Articles in the publication
Clinical Physiol ...
By the university
Örebro University
Linköping University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view