On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry

Andreo, Pedro (author)
Karolinska Institutet,Stockholms universitet,Fysikum
Burns, David T. (author)
Salvat, Francesc (author)
 (creator_code:org_t)
2012-03-27
2012
English.
In: Physics in Medicine and Biology. - : IOP Publishing. - 0031-9155 .- 1361-6560. ; 57:8, s. 2117-2136
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • A systematic analysis of the available data has been carried out for mass energy-absorption coefficients and their ratios for air, graphite and water for photon energies between 1 keV and 2 MeV, using representative kilovoltage x-ray spectra for mammography and diagnostic radiology below 100 kV, and for Ir-192 and Co-60 gamma-ray spectra. The aim of this work was to establish 'an envelope of uncertainty' based on the spread of the available data. Type A uncertainties were determined from the results of Monte Carlo (MC) calculations with the PENELOPE and EGSnrc systems, yielding mean values for mu(en)/rho with a given statistical standard uncertainty. Type B estimates were based on two groupings. The first grouping consisted of MC calculations based on a similar implementation but using different data and/or approximations. The second grouping was formed by various datasets, obtained by different authors or methods using the same or different basic data, and with different implementations (analytical, MC-based, or a combination of the two); these datasets were the compilations of NIST, Hubbell, Johns-Cunningham, Attix and Higgins, plus MC calculations with PENELOPE and EGSnrc. The combined standard uncertainty, u(c), for the mu(en)/rho values for the mammography x-ray spectra is 2.5%, decreasing gradually to 1.6% for kilovoltage x-ray spectra up to 100 kV. For Co-60 and Ir-192, u(c) is approximately 0.1%. The Type B uncertainty analysis for the ratios of mu(en)/rho values includes four methods of analysis and concludes that for the present data the assumption that the data interval represents 95% confidence limits is a good compromise. For the mammography x-ray spectra, the combined standard uncertainties of (mu(en)/rho)(graphite,air) and (mu(en)/rho)(graphite,water) are 1.5%, and 0.5% for (mu(en)/rho)(water,air), decreasing gradually down to u(c) = 0.1% for the three mu(en)/rho ratios for the gamma-ray spectra. The present estimates are shown to coincide well with those of Hubbell (1977 Rad. Res. 70 58-81), except for the lowest energy range (radiodiagnostic) where it is concluded that current databases and their systematic analysis represent an improvement over the older Hubbell estimations. The results for (mu(en)/rho)(graphite,air) for the gamma-ray dosimetry range are moderately higher than those of Seltzer and Bergstrom (2005 private communication).

Subject headings

NATURVETENSKAP  -- Fysik (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences (hsv//eng)

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Andreo, Pedro
Burns, David T.
Salvat, Francesc
About the subject
NATURAL SCIENCES
NATURAL SCIENCES
and Physical Science ...
Articles in the publication
Physics in Medic ...
By the university
Stockholm University
Karolinska Institutet

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view