Dopamine D2 receptor SPECT with (123)I-IBZM : evaluation of collimator and post-filtering when using model-based compensation-a Monte Carlo study.

Larsson, Anne (author)
Umeå universitet,Radiofysik
Jakobson Mo, Susanna (author)
Umeå universitet,Diagnostisk radiologi
Ljungberg, Michael (author)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
show more...
Riklund, Katrine (author)
Umeå universitet,Diagnostisk radiologi
show less...
 (creator_code:org_t)
2010-03-12
2010
English.
In: Physics in Medicine and Biology. - : IOP Publishing. - 0031-9155 .- 1361-6560. ; 55:7, s. 1971-1988
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • In (123)I-IBZM brain SPECT, the main interest is the activity uptake in the striatum relative to the background, and semi-quantitative techniques using regions of interest are typically used for this purpose. Uncertainties in the measured uptakes can however be a problem due to low contrasts and high noise levels. Like SPECT in general, IBZM SPECT should benefit from reconstruction methods that include model-based compensation, but it is important that image acquisition is optimized for this technique. An important factor is the choice of collimator. In this study we compare four different parallel-hole collimators for IBZM SPECT regarding overall quantitative accuracy and measured uptake ratio as a function of image noise and uncertainty. The collimators are low-energy high-resolution (LEHR), low-energy general-purpose (LEGP), extended LEGP (ELEGP) and medium-energy general-purpose (MEGP). The effect of three Butterworth post-filters with cut-off frequencies of 0.3, 0.45 and 0.6 cm(-1) (power factor 8) is also studied. All raw-data projections are produced using Monte Carlo simulations. Of the investigated collimators, the one that is most sensitive to the primary photons, ELEGP, proved to be the most optimal for realistic noise levels. Butterworth post-filtering is advantageous, and the cut-off frequency 0.45 cm(-1) was the best compromise in this study.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Larsson, Anne
Jakobson Mo, Sus ...
Ljungberg, Micha ...
Riklund, Katrine
About the subject
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Clinical Medicin ...
and Radiology Nuclea ...
Articles in the publication
Physics in Medic ...
By the university
Umeå University
Lund University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view