Variability in Heart-to-Mediastinum Ratio from Planar 123I-MIBG Images of a Thorax Phantom for 6 Common γ-Camera Models

Owenius, Rikard (author)
Uppsala universitet,Radiologi
Zanette, Michelle (author)
GE Healthcare, Life Sci, Imaging R&D, Marlborough, MA USA.
Cella, Patrick (author)
GE Healthcare, Life Sci, Imaging R&D, Marlborough, MA USA.
 (creator_code:org_t)
2017-10-17
2017
English.
In: Journal of Nuclear Medicine Technology. - : Society of Nuclear Medicine. - 0091-4916 .- 1535-5675. ; 45:4, s. 297-303
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • A heart-to-mediastinum (H/M) ratio of 1.6 or greater on planar I-123-iobenguane (I-123-MIBG) images identifies heart failure patients at low risk of experiencing an adverse cardiac event. This phase-4 study used standardized phantoms to assess the intercamera, intracamera, and interhead variability in H/M ratio determinations from planar cardiac I-123-MIBG imaging using commercially available, dual-head.-cameras. Methods: A fillable thorax phantom was developed to simulate the typical uptake of I-123-MIBG. The phantom had a nominal H/M ratio of 1.6 on the reference camera. Commercial cameras used in the study were dual-head and capable of 90 degrees configuration for cardiac imaging. The target sample size was 8 units (examples) per camera model. Two imaging technologists independently analyzed planar images of simulated I-123-MIBG uptake from the thorax phantom. H/M was the ratio of the average counts per pixel of the heart and mediastinum regions of interest. The primary endpoint, intercamera variability in H/M ratio from head 1, was determined for each camera model via comparison with the H/M ratio on the reference camera. Only cameras with at least 8 units tested (n >= 8) were included in the primary analysis. Intracamera and interhead variability in the H/M ratio were also evaluated. Results: Nine camera models were studied. The mean H/M ratio ranged from 1.342 to 1.677. The primary analysis (6 camera models) using a mixed-model, repeated-measures analysis showed no significant difference in H/M ratio between any camera model and the reference camera. Intracamera variability (head 1) in the H/M ratio among camera models with 8 units or more was high, with SDs ranging from 0.0455 to 0.1193. Interhead variability was low (SDs of the interhead difference, 0.017-0.074). Conclusion: Commonly used.-cameras produced H/M ratios from simulated I-123-MIBG phantom images that were not significantly different from those on the reference camera. This finding indicates that the results of previous clinical trials of I-123-MIBG, involving many different clinical sites and camera models, are valid. The assessment of the performance of a given camera unit using an I-123 planar phantom before H/M results from I-123-MIBG imaging are used for classifying risk in heart failure patients is encouraged.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

iobenguane
MIBG
phantom
heart-to-mediastinum ratio
heart failure

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Owenius, Rikard
Zanette, Michell ...
Cella, Patrick
About the subject
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Clinical Medicin ...
and Radiology Nuclea ...
Articles in the publication
Journal of Nucle ...
By the university
Uppsala University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view