Pulmonary nodule size evaluation with chest tomosynthesis.

Johnsson, Åse (Allansdotter), 1966 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Fagman, Erika (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Vikgren, Jenny, 1957 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
show more...
Fisichella, Valeria A, 1974 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Boijsen, Marianne, 1948 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Flinck, Agneta, 1953 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Kheddache, Susanne, 1949 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiologi,Institute of Clinical Sciences, Department of Radiology
Svalkvist, Angelica (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Båth, Magnus, 1974 (author)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
show less...
 (creator_code:org_t)
Radiological Society of North America (RSNA), 2012
2012
English.
In: Radiology. - : Radiological Society of North America (RSNA). - 1527-1315 .- 0033-8419. ; 265:1, s. 273-82
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Purpose: To evaluate intra- and interobserver variability, as well as agreement for nodule size measurements on chest tomosynthesis and computed tomographic (CT) images. Materials and Methods: The Regional Ethical Review Board approved this study, and all participants gave written informed consent. Thirty-six segmented nodules in 20 patients were included in the study. Eight observers measured the left-to-right, inferior-to-superior, and longest nodule diameters on chest tomosynthesis and CT images. Intra- and interobserver repeatability, as well as agreement between measurements on chest tomosynthesis and CT images, were assessed as recommended by Bland and Altman. Results: The difference between the mean manual and the segmented diameter was -2.2 and -2.3 mm for left-to-right and -2.6 and -2.2 mm for the inferior-to-superior diameter for measurements on chest tomosynthesis and CT images, respectively. Intraobserver 95% limits of agreement (LOA) for the longest diameter ranged from a lower limit of -1.1 mm and an upper limit of 1.0 mm to -1.8 and 1.8 mm for chest tomosynthesis and from -0.6 and 0.9 mm to -3.1 and 2.2 mm for axial CT. Interobserver 95% LOA ranged from -1.3 and 1.5 mm to -2.0 and 2.1 mm for chest tomosynthesis and from -1.8 and 1.1 mm to -2.2 and 3.1 mm for axial CT. The 95% LOA concerning the mean of the observers' measurements of the longest diameter at chest tomosynthesis and axial CT were ±2.1 mm (mean measurement error, 0 mm). For the different observers, the 95% LOA between the modalities ranged from -2.2 and 1.6 mm to -3.2 and 2.8 mm. Conclusion: Measurements on chest tomosynthesis and CT images are comparable, because there is no evident bias between the modalities and the repeatability is similar. The LOA between measurements for the two modalities raise concern if measurements from chest tomosynthesis and CT were to be used interchangeably. © RSNA, 2012 Supplemental material: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.12111459/-/DC1.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

  • Radiology (Search for host publication in LIBRIS)

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view