Search: onr:"swepub:oai:gup.ub.gu.se/282794" >
A sensitive method ...
A sensitive method detecting trace levels of levonorgestrel using LC-HRMS
-
- Abujrais, Sandy (author)
- Uppsala universitet,Klinisk kemi
-
- Olovsson, Matts, 1958- (author)
- Uppsala universitet,Reproduktionsbiologi
-
- Ahnoff, Martin (author)
- Gothenburg University,Göteborgs universitet,Institutionen för marina vetenskaper,Department of marine sciences,Department of Marine Sciences, University of Gothenburg, Carl Skottbergs gata 22B, SE-41319 Gothenburg, Sweden.
-
show more...
-
- Rasmusson, Annica (author)
- Uppsala universitet,Klinisk farmakogenomik och osteoporos
-
- Larsson, Anders (author)
- Uppsala universitet,Klinisk kemi
-
- Åkerfeldt, Torbjörn (author)
- Uppsala universitet,Klinisk kemi
-
- Kultima, Kim (author)
- Uppsala universitet,Klinisk kemi
-
show less...
-
(creator_code:org_t)
- Elsevier BV, 2019
- 2019
- English.
-
In: Contraception. - : Elsevier BV. - 0010-7824 .- 1879-0518. ; 100:3, s. 247-249
- Related links:
-
https://gup.ub.gu.se...
-
show more...
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
show less...
Abstract
Subject headings
Close
- Objective: To develop a high resolution mass spectrometry (HRMS) method to quantify levonorgestrel (LNG) in serum. Study design: Levonorgestrel was extracted using solid phase extraction and measured using liquid chromatography (LC) HRMS. Results: Low limit of quantification (LLOQ) was 25 pg/mL and low limit of detection (LLOD) was 12.5 pg/mL. Precision and accuracy bias were <10%. LNG in serum samples from Mirena® users ranged between 37 to 219 pg/mL (n=12). In eight out of 22 patients with suspected intrauterine device (IUD) expulsion LNG was detected (26–1272 pg/mL). Conclusion: A sensitive, fast and simple LC-HRMS method was developed to detect trace levels of LNG. © 2019 Elsevier Inc.
Subject headings
- NATURVETENSKAP -- Biologi -- Biokemi och molekylärbiologi (hsv//swe)
- NATURAL SCIENCES -- Biological Sciences -- Biochemistry and Molecular Biology (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Reproduktionsmedicin och gynekologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Obstetrics, Gynaecology and Reproductive Medicine (hsv//eng)