Effect of Precision on Longitudinal Follow-Up of Bone Mineral Density Measurements in Elderly Women and Men.
-
Lenora, Janaka (author)
-
- Åkesson, Kristina (author)
- Lund University,Lunds universitet,Ortopedi - klinisk och molekylär osteoporosforskning,Forskargrupper vid Lunds universitet,Orthopedics - Clinical and Molecular Osteoporosis Research,Lund University Research Groups
-
- Gerdhem, Paul (author)
- Lund University,Lunds universitet,Ortopedi - klinisk och molekylär osteoporosforskning,Forskargrupper vid Lunds universitet,Orthopedics - Clinical and Molecular Osteoporosis Research,Lund University Research Groups
-
(creator_code:org_t)
- Elsevier BV, 2010
- 2010
- English.
-
In: Journal of Clinical Densitometry. - : Elsevier BV. - 1094-6950. ; 13:4, s. 407-412
- Related links:
-
https://portal.resea... (primary) (free)
-
show more...
-
http://www.ncbi.nlm....
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
http://kipublication...
-
show less...
Abstract
Subject headings
Close
- Precision error of dual-energy X-ray absorptiometry exceeds the expected annual rate of bone loss in the elderly. The capacity to detect changes in areal bone mineral density (aBMD; g/cm(2)) over a 5-yr period was assessed. Six hundred ninety-one women, 75.2 (0.1)yr, from the Malmö OPRA-study, were measured using Lunar DPX-L (GE Lunar, Madison, WI), and 211 men, 74.7 (3.2)yr, from the Malmö Mr Os-study, were measured using Lunar Prodigy (GE Lunar) with follow-up 5yr later. Precision error was determined with 30 degrees of freedom. Least significant change (LSC, i.e., 2.77xprecision error) was calculated. Women's precision errors (g/cm(2)) for DPX-L were 0.028 (total hip [TH]) and 0.016 (lumbar spine [LS]), and for Prodigy, they were 0.009 (TH) and 0.039 (LS). In men, corresponding results for Prodigy were 0.014 and 0.031. In women, 41% and in men, 39% had aBMD changes exceeding the LSC at TH. Follow-up intervals (i.e., LSC/median rate of aBMD change) for both women and men were 8yr (TH) and 13yr (LS). Based on Prodigy precision data, follow-up intervals for women were 3 and 32yr at TH and LS. In summary, several years were needed to detect change. Only when a high rate of bone loss is suspected, a short follow-up time is possible, in elderly persons.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Ortopedi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Orthopaedics (hsv//eng)
Keyword
- Duel energy x-ray absorptiometry
- longitudinal follow-up
- least significant change
- elderly individuals
- precision error.