SwePub
Sök i LIBRIS databas

  Extended search

onr:"swepub:oai:lup.lub.lu.se:57706479-a119-4364-9e91-2d98281ef546"
 

Search: onr:"swepub:oai:lup.lub.lu.se:57706479-a119-4364-9e91-2d98281ef546" > Time-frequency char...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Time-frequency characterization of simultaneous intra-atrial and electrocardiographic recordings during atrial fibrillation

Stridh, Martin (author)
Lund University,Lunds universitet,Institutionen för elektro- och informationsteknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Electrical and Information Technology,Departments at LTH,Faculty of Engineering, LTH
Bollmann, Andreas (author)
Husser, D (author)
show more...
Bhandari, A (author)
Cannom, D (author)
Sörnmo, Leif (author)
Lund University,Lunds universitet,Institutionen för elektro- och informationsteknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Electrical and Information Technology,Departments at LTH,Faculty of Engineering, LTH
show less...
 (creator_code:org_t)
2005
2005
English 4 s.
In: Computers in Cardiology 2005. - 0780393376 ; 32, s. 347-350
  • Conference paper (peer-reviewed)
Abstract Subject headings
Close  
  • A new method for characterization of simultaneous intra-atrial and ECG recordings during atrial fibrillation (AF) is presented. With different preprocessing, both types of signals are characterized using a logarithmic time-frequency distribution from which trends of frequency, regularity, morphology (only ECG) and signal quality are extracted. The objective of the study is to relate ECG measures of rate and morphology to measures of intra-atrial organisation. The algorithm has been tested on a database with simultaneous ECG and right atrial recordings from 34 patients with drug-refractory AF. The average frequency in lead V1 was 5.91 Hz (std=0.94) and the average exponential decay of harmonic magnitudes (reflecting the morphology) in V1 was 1.31 (std=0.38); their correlation coefficient was 0.53 (p<0.001). The correlation coefficient between spatial frequency dispersion in the right atrium and harmonic decay was 0.53 (p<0.005). More organized AF, observed in patients treated with amiodarone, was reflected by a lower frequency, smaller right atrial frequency dispersion and lower exponential decay

Subject headings

TEKNIK OCH TEKNOLOGIER  -- Elektroteknik och elektronik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Electrical Engineering, Electronic Engineering, Information Engineering (hsv//eng)

Keyword

harmonic decay
right atrium
electrocardiography
intra-atrial recordings
atrial fibrillation
ECG
spatial frequency dispersion
logarithmic time-frequency distribution
correlation coefficient
drug-refractory AF

Publication and Content Type

kon (subject category)
ref (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Stridh, Martin
Bollmann, Andrea ...
Husser, D
Bhandari, A
Cannom, D
Sörnmo, Leif
About the subject
ENGINEERING AND TECHNOLOGY
ENGINEERING AND ...
and Electrical Engin ...
Articles in the publication
Computers in Car ...
By the university
Lund University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view