SwePub
Sök i LIBRIS databas

  Extended search

onr:"swepub:oai:lup.lub.lu.se:6bed5544-e8eb-4972-9923-9ee970d8a2ec"
 

Search: onr:"swepub:oai:lup.lub.lu.se:6bed5544-e8eb-4972-9923-9ee970d8a2ec" > Comparison between ...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Comparison between tape-recorded and amplitude-integrated EEG monitoring in sick newborn infants

Hellström-Westas, Lena (author)
Lund University,Lunds universitet,Pediatrik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Paediatrics (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
 (creator_code:org_t)
Wiley, 1992
1992
English.
In: Acta Pædiatrica. - : Wiley. - 1651-2227 .- 0803-5253. ; 81:10, s. 812-819
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • In 15 ill newborn infants a comparison between long-term multichannel and single-channel recordings of simultaneously tape-recorded (Medilog system) and amplitude-integrated EEG (Cerebral Function Monitor) was made. There was good agreement between the main type of background activity diagnosed with the tape-recorded and the amplitude-integrated EEG for all recordings. Two infants had repetitive subclinical and subtle seizure activity, lasting for several hours, which was detected by both techniques. Short, single seizures were diagnosed in the recordings of nine infants. When a single electrographic seizure appeared in an otherwise stable recording, it was identified by both the tape-recorded and the amplitude-integrated EEG. Very short (5-30 s) seizure patterns, which were diagnosed with the tape-recorded EEG, were not identified in the cerebral function monitor recordings. In the single-channel recordings of both the EEG and the cerebral function monitor there were, on some occasions, difficulties in distinguishing single seizures from interference due to external artefacts. In the multichannel recordings the diagnosis of seizure patterns was facilitated by comparison with the other channels. Both the Medilog EEG and the cerebral function monitor are feasible techniques for following cerebral electrical activity in sick neonates, although neither technique is specifically constructed for this purpose. For clinical use in the neonatal intensive care unit the advantage with the cerebral function monitor is the immediately available recording. The tape-recorded EEG offers possibilities of more channels and a higher reliability when diagnosing short subclinical seizures, however, only after offline analysis.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Pediatrik (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Pediatrics (hsv//eng)

Keyword

Amplitude-integrated EEG
cerebral function monitor
EEG monitoring
neonates
tape-recorded EEG

Publication and Content Type

art (subject category)
ref (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Find more in SwePub

By the author/editor
Hellström-Westas ...
About the subject
MEDICAL AND HEALTH SCIENCES
MEDICAL AND HEAL ...
and Clinical Medicin ...
and Pediatrics
Articles in the publication
Acta Pædiatrica
By the university
Lund University

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view