Endovascular AAA exclusion: will stents with hooks and barbs prevent stent-graft migration?
-
- Malina, Martin (author)
- Lund University,Lunds universitet,Institutionen för kliniska vetenskaper, Malmö,Medicinska fakulteten,Vaskulära sjukdomar - kliniska studier,Forskargrupper vid Lunds universitet,Department of Clinical Sciences, Malmö,Faculty of Medicine,Vascular Diseases - Clinical Research,Lund University Research Groups
-
- Lindblad, Bengt (author)
- Lund University,Lunds universitet,Vaskulära sjukdomar - kliniska studier,Forskargrupper vid Lunds universitet,Vascular Diseases - Clinical Research,Lund University Research Groups
-
- Ivancev, Krassi (author)
- Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
-
show more...
-
- Lindh, Mats (author)
- Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
-
Malina, Janne (author)
-
Brunkwall, Jan (author)
-
show less...
-
(creator_code:org_t)
- 1998
- 1998
- English.
-
In: Journal of Endovascular Surgery. - 1074-6218. ; 5:4, s. 310-317
- Related links:
-
http://dx.doi.org/10...
-
show more...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- PURPOSE: To investigate if stents with hooks and barbs will improve stent-graft fixation in the abdominal aorta. METHODS: Sixteen- to 24-mm-diameter Dacron grafts were deployed inside cadaveric aortas. The grafts were anchored by stents as in endovascular abdominal aortic aneurysm repair. One hundred thirty-seven stent-graft deployments were carried out with modified self-expanding Z-stents with (A) no hooks and barbs (n = 75), (B) 4 5-mm-long hooks and barbs (n = 39), (C) 8 10-mm-long, strengthened hooks and barbs (n = 19), or (D) hooks only (n = 4). Increasing longitudinal traction was applied to determine the displacement force needed to extract the stent-grafts. The radial force of the stents was measured and correlated to the displacement force. RESULTS: The median (interquartile range) displacement force needed to extract grafts anchored by stent A was 2.5 N (2.0 to 3.4), stent B 7.8 N (7.4 to 10.8), and stent C 22.5 N (17.1 to 27.9), p < 0.001. Both hooks and barbs added anchoring strength. During traction, the weaker barbs were distorted or caused intimal tears. The stronger barbs engaged the entire aortic wall. The radial force of the stents had no impact on fixation, while aortic calcification and graft oversizing had marginal effects. CONCLUSIONS: Stent barbs and hooks increased the fixation of stent-grafts tenfold, while the radial force of stents had no impact. These data may prove important in future endograft development to prevent stent-graft migration after aneurysm exclusion.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Kardiologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)
Keyword
- endovascular grafting
- abdominal aortic aneurysm
- migration
- Z-stents