Investigating photomultiplier tube nonlinearities in high-speed phosphor thermometry using light emitting diode simulated decay curves
-
- Feuk, Henrik (author)
- Lund University,Lunds universitet,Förbränningsfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,LTH profilområde: Energiomställningen,LTH profilområden,Combustion Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH,LTH Profile Area: The Energy Transition,LTH Profile areas,Faculty of Engineering, LTH
-
- Nilsson, Sebastian (author)
- Lund University,Lunds universitet,Förbränningsfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,LTH profilområde: Energiomställningen,LTH profilområden,LTH profilområde: Teknik för hälsa,Combustion Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH,LTH Profile Area: The Energy Transition,LTH Profile areas,Faculty of Engineering, LTH,LTH Profile Area: Engineering Health,Faculty of Engineering, LTH
-
- Aldén, Marcus (author)
- Lund University,Lunds universitet,Förbränningsfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,LTH profilområde: Energiomställningen,LTH profilområden,Combustion Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH,LTH Profile Area: The Energy Transition,LTH Profile areas,Faculty of Engineering, LTH
-
show more...
-
- Richter, Mattias (author)
- Lund University,Lunds universitet,Förbränningsfysik,Fysiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,LTH profilområde: Avancerade ljuskällor,LTH profilområden,LTH profilområde: Energiomställningen,Combustion Physics,Department of Physics,Departments at LTH,Faculty of Engineering, LTH,LTH Profile Area: Photon Science and Technology,LTH Profile areas,Faculty of Engineering, LTH,LTH Profile Area: The Energy Transition,Faculty of Engineering, LTH
-
show less...
-
(creator_code:org_t)
- AIP Publishing, 2021
- 2021
- English 11 s.
-
In: Review of Scientific Instruments. - : AIP Publishing. - 1089-7623 .- 0034-6748. ; 92:12
- Related links:
-
http://dx.doi.org/10... (free)
-
show more...
-
https://aip.scitatio...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- Photomultiplier tube (PMT) nonlinearities relevant for single shot high-speed lifetime phosphor thermometry were investigated by simulating decay curves with a light emitting diode (LED) at repetition rates between 1 Hz and 10 kHz. The PMT gain, LED decay time, and background radiant flux were also varied to investigate their impact on the measured decay time error. Errors in the measured decay time due to nonlinear PMT performance lead to temperature measurement errors; therefore, having the measured decay time sensitive to only phosphor temperature is highly valuable for more reliable temperature measurements. Photocathode bleaching had a significant impact on the signal level linearity for PMTs with excitation frequency in the kHz regime but had a smaller impact on the decay time error. Space charge effects were most noticeable at high radiant flux levels and high repetition rates. Strong background radiant flux may lead to decay time errors, and a gateable photocathode could be an effective method to reduce decay time errors. The best decay time measurement configuration to maximize precision without sacrificing accuracy is to use PMT gain in the recommended range and the highest radiant flux where the PMT response is still linear. The degree of nonlinearity in the PMT response is partly detector dependent; therefore, the results in this work may differ among detectors; however, the analysis presented in this work provides guidelines for improving the temperature accuracy of kHz lifetime phosphor thermometry measurements.
- Photomultiplier tube (PMT) nonlinearities relevant for single shot high-speed lifetime phosphor thermometry were investigated by simulating decay curves with a light emitting diode (LED) at repetition rates between 1 Hz and 10 kHz. The PMT gain, LED decay time, and background radiant flux were also varied to investigate their impact on the measured decay time error. Errors in the measured decay time due to nonlinear PMT performance lead to temperature measurement errors; therefore, having the measured decay time sensitive to only phosphor temperature is highly valuable for more reliable temperature measurements. Photocathode bleaching had a significant impact on the signal level linearity for PMTs with excitation frequency in the kHz regime but had a smaller impact on the decay time error. Space charge effects were most noticeable at high radiant flux levels and high repetition rates. Strong background radiant flux may lead to decay time errors, and a gateable photocathode could be an effective method to reduce decay time errors. The best decay time measurement configuration to maximize precision without sacrificing accuracy is to use PMT gain in the recommended range and the highest radiant flux where the PMT response is still linear. The degree of nonlinearity in the PMT response is partly detector dependent; therefore, the results in this work may differ among detectors; however, the analysis presented in this work provides guidelines for improving the temperature accuracy of kHz lifetime phosphor thermometry measurements.
Subject headings
- NATURVETENSKAP -- Fysik -- Atom- och molekylfysik och optik (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Atom and Molecular Physics and Optics (hsv//eng)
Publication and Content Type
- art (subject category)
- ref (subject category)
Find in a library
To the university's database