Search: onr:"swepub:oai:lup.lub.lu.se:b6538931-6b02-46a6-b87d-18ceb67ef34b" >
Multifocal visual e...
Multifocal visual evoked potentials-a method study of responses from small sectors of the visual field.
-
- Bengtsson Lindberg, Marie (author)
- Lund University,Lunds universitet,Klinisk neurofysiologi,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Neurophysiology,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Andréasson, Sten (author)
- Lund University,Lunds universitet,Oftalmologi, Lund,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Ophthalmology, Lund,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Andersson, Gert (author)
- Lund University,Lunds universitet,Klinisk neurofysiologi,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Clinical Neurophysiology,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine
-
(creator_code:org_t)
- Elsevier BV, 2005
- 2005
- English.
-
In: Clinical Neurophysiology. - : Elsevier BV. - 1872-8952 .- 1388-2457. ; 116:8, s. 1975-1983
- Related links:
-
http://www.ncbi.nlm....
-
show more...
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- Objective: A method study of the mfVEP technique to establish a standardised way to identify stable response components from small areas in all parts of the visual field and a test-retest reliability study. Methods: MfVEP was recorded from 26 healthy volunteers. Results: Two response components could be clearly identified. The latencies corresponded to those of the traditional VEP response (N75 and P100). The visual field was divided into 12 sectors. A characteristic pattern was obtained. Component I was mainly negative in the upper sectors and positive in the lower sectors. Component II was positive in the upper sectors and negative in the lower ones. Most of the sectors with missing responses were the ones adjacent to the horizontal meridian, corresponding to the phase reversals. In a test-retest reliability study, the amplitude and latency measurements of the second test were plotted against those of the first test. Correlation coefficients between 0.84 and 0.93 were obtained. Conclusions: The mfVEP allows a reliable quantification of two response components from small parts of the visual field. Significance: This paper suggests that mfVEP could be a valuable supplement to the traditional VEP for exploring restricted parts of the visual pathways.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Neurologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Neurology (hsv//eng)
Keyword
- reproducibility
- normal responses
- VEP
- multifocal visual evoked potential (MtVEP)
Publication and Content Type
- art (subject category)
- ref (subject category)
Find in a library
To the university's database