Search: onr:"swepub:oai:lup.lub.lu.se:b8634e4d-befa-42aa-b062-2e3ce6d04138" >
Biosensor based on ...
Biosensor based on cellobiose dehydrogenase for detection of catecholamines
-
- Stoica, Leonard (author)
- Lund University,Lunds universitet,Centrum för analys och syntes,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Centre for Analysis and Synthesis,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
-
Lindgren-Sjolander, A (author)
-
- Ruzgas, Tautgirdas (author)
- Lund University,Lunds universitet,Centrum för analys och syntes,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Centre for Analysis and Synthesis,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
-
show more...
-
- Gorton, Lo (author)
- Lund University,Lunds universitet,Centrum för analys och syntes,Kemiska institutionen,Institutioner vid LTH,Lunds Tekniska Högskola,Centre for Analysis and Synthesis,Department of Chemistry,Departments at LTH,Faculty of Engineering, LTH
-
show less...
-
(creator_code:org_t)
- 2004-06-29
- 2004
- English.
-
In: Analytical Chemistry. - : American Chemical Society (ACS). - 1520-6882 .- 0003-2700. ; 76:16, s. 4690-4696
- Related links:
-
http://dx.doi.org/10...
-
show more...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- A cellobiose dehydrogenase (CDH)-modified graphite electrode was designed for amperometric detection of catecholamines in the flow injection mode, by their recycling between the graphite electrode (+300 mV vs AgAgCl) and the reduced FAD cofactor of adsorbed CDH, resulting in an amplified response signal. The high efficiency of the enzyme-catecholamine reaction leads to a detection limit below 1 nM and a sensitivity of 15.8 A(.)M(-1.)cm(2) 2 (1150 nA/muM) for noradrenaline, with a coverage of less than 2.5 mug of CDH adsorbed on the electrode surface (0.073 cm(2)). Working parameters such as pH, cellobiose concentration, carrier buffer, and applied potential were optimized, using hydroquinone as a model analyte. The sensitivity, linear range, and amplification factor can be modulated by the steady-state concentration of cellobiose in the flow buffer. The response of the sensor decreases only 2% when run continuously for 4 h in the flow injection mode. The response peak maximum is obtained within 6 s at a flow rate of 0.5 mL/min, representing the time of the entire sample segment to pass the electrode. CDH enzymes from Phanerochaete chrysosporium and Sclerotium rolfsii were investigated, providing different characteristics of the sensor, with sensors made with CDH from P. chrysosporium being the better ones.
Subject headings
- NATURVETENSKAP -- Kemi -- Analytisk kemi (hsv//swe)
- NATURAL SCIENCES -- Chemical Sciences -- Analytical Chemistry (hsv//eng)
Publication and Content Type
- art (subject category)
- ref (subject category)
Find in a library
To the university's database