Search: onr:"swepub:oai:lup.lub.lu.se:cc0fea6c-97f8-4834-97c8-f1507054c799" >
Beam commissioning ...
Beam commissioning and measurements validating the beam model in a new TPS that converts helical tomotherapy plans to step-and-shoot IMRT plans.
-
- Petersson, Kristoffer (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Ceberg, Crister (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Engström, Per (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
show more...
-
- Knöös, Tommy (author)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
show less...
-
(creator_code:org_t)
- 2010-12-14
- 2011
- English.
-
In: Medical Physics. - : Wiley. - 0094-2405. ; 38:1, s. 40-46
- Related links:
-
https://portal.resea... (primary) (free)
-
show more...
-
http://www.ncbi.nlm....
-
http://dx.doi.org/10...
-
https://lup.lub.lu.s...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- A new type of treatment planning system called SHAREPLAN has been studied, which enables the transfer of treatment plans generated for helical tomotherapy delivery to plans that can be delivered on C-arm linacs. The purpose is to ensure continuous patient treatment during periods of unscheduled downtime for the TomoTherapy unit, particularly in clinics without a backup unit. The purpose of this work was to verify that the plans generated in this novel planning system are deliverable and accurate. The work consists primarily of beam commissioning, verification of the beam model, and measurements verifying that generated plans are deliverable with sufficient accuracy.
Subject headings
- NATURVETENSKAP -- Biologi -- Biofysik (hsv//swe)
- NATURAL SCIENCES -- Biological Sciences -- Biophysics (hsv//eng)
Keyword
- Computer-Assisted: methods
- Radiotherapy Planning
- Intensity-Modulated: methods
- Radiotherapy
Publication and Content Type
- art (subject category)
- ref (subject category)
Find in a library
To the university's database