Evaluation of an external device measuring knee joint rotation: an in vivo study with simultaneous Roentgen stereometric analysis.
-
- Almquist, Per Otto (author)
- Lund University,Lunds universitet,Människan i rörelse: hälsa och rehabilitering,Forskargrupper vid Lunds universitet,Human Movement: health and rehabilitation,Lund University Research Groups
-
Arnbjörnsson, Arnbjörn (author)
-
Zätterström, Rose (author)
-
show more...
-
- Ryd, Leif (author)
- Lund University,Lunds universitet,Ortopedi, Lund,Sektion III,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Orthopaedics (Lund),Section III,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Ekdahl, Charlotte (author)
- Lund University,Lunds universitet,Människan i rörelse: hälsa och rehabilitering,Forskargrupper vid Lunds universitet,Human Movement: health and rehabilitation,Lund University Research Groups
-
- Fridén, Thomas (author)
- Lund University,Lunds universitet,Ortopedi, Lund,Sektion III,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Orthopaedics (Lund),Section III,Department of Clinical Sciences, Lund,Faculty of Medicine
-
show less...
-
(creator_code:org_t)
- 2002
- 2002
- English.
-
In: Journal of Orthopaedic Research. - 1554-527X. ; 20:3, s. 427-432
- Related links:
-
http://dx.doi.org/10...
-
show more...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
show less...
Abstract
Subject headings
Close
- An external device ("rottometer") specially designed to measure knee joint rotation was developed and evaluated with respect to its validity. Simultaneous measurements were made with the rottometer and Roentgen stereometric analysis (RSA) in five patients with implanted tantalum markers in the tibia and femur. Measurements of internal and external rotation were made at 90 degrees and 60 degrees of knee flexion using 3, 6 and 9 N m torques. The coefficients of determination (r2) between the results obtained with the rottometer and RSA were around 0.9 for the total rotation. The rottometer consistently overestimated the rotation by about 100% and this systematic error was most constant at 90 degrees flexion for the different torques. The magnitude of this error from soft tissue deformation as well as the rotatory movements in the hip, foot and ankle joints must be considered when using external devices to measure knee rotation in clinical studies. The most accurate registrations were found in 90 degrees flexion with 9 N m force (r2 = 0.94).
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Ortopedi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Orthopaedics (hsv//eng)
Keyword
- Arthrography
- Equipment and Supplies
- Human
- Knee Joint : physiopathology
- Knee Joint : radiography
- Range of Motion
- Male
- Articular
- Rotation
- Equipment Design