Characterization of HPGe detectors using Computed Tomography

Hedman, Angelica, 1984- (author)
Umeå universitet,Radiofysik,Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
Gogani, J. B. (author)
Totalförsvarets forskningsinstitut (FOI),Swedish Defence Research Agency (FOI),Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
Granstrom, M. (author)
Totalförsvarets forskningsinstitut (FOI),Swedish Defence Research Agency (FOI),Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden
show more...
Johansson, Lennart (author)
Umeå universitet,Radiofysik
Andersson, Jonas S. (author)
Umeå universitet,Institutionen för strålningsvetenskaper
Ramebäck, Henrik, 1967 (author)
Chalmers tekniska högskola,Chalmers University of Technology,Swedish Defence Research Agency, Division of CBRN Defence and Security, SE-90182 Umeå, Sweden; Chalmers University of Technology, Department of Chemical and Biological Engineering, Nuclear Chemistry, SE-41296 Göteborg, Sweden
show less...
 (creator_code:org_t)
Elsevier BV, 2015
2015
English.
In: Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. - : Elsevier BV. - 0168-9002 .- 1872-9576. ; 785, s. 21-25
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Computed Tomography (CT) high resolution imaging have been used to investigate if there is a significant change in the crystal-to-window distance, i.e. the air gap thickness, in a small n-type detector cooled to 77 K, and in a medium sized p-type HPGe detector when cooled to 100 K. The findings were compared to detector dimension data made available by the manufacturer. The air gap thickness increased by (0.38 +/- 0.07) mm for the n-type detector and by (0.40 +/- 0.15) mm for the p-type detector when the detectors were cooled to 77 resp. 100 K compared to at room temperature. Monte Carlo calculations indicate that these differences have a significant impact on the efficiency in close geometries (

Subject headings

NATURVETENSKAP  -- Fysik -- Subatomär fysik (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences -- Subatomic Physics (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Keyword

Semi-empirical calibration
Monte Carlo modelling
Gamma spectrometry
Gespecor
Computed Tomography
HPGe

Publication and Content Type

art (subject category)
ref (subject category)

Find in a library

To the university's database

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Search outside SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view